Calibration (CAL):

Both zero and span adjustments are made 
through the menu.  (For complete calibration 
details of the GXLdp without using the Spool 
Cal option, refer to the RECAL section- page 8).
A 90-degree clockwise rotation of the 
SpoolCal key places the unit into ‘CAL’ mode, 
this isolates the GXLdp from the process 
pressure.
The RECAL function in the GXLdp menu pro-
vides the user the ability to field calibrate the 
product.  The most recent factory calibration 
is permanently retained in memory and can 
be recalled at any time.

Once you enter the Re-Calibration (RECAL) 
Mode you will have the following options:
 1. 

FACT

 – Factory Calibration

 2. 

NEW

 – New Field Calibration

 3. 

FIELD

 – Field Calibration 

Note:

 On initial calibration, Field Calibration 

will not show in the menu. To perform the ini-
tial field calibration you will choose New (New 
Field Calibration). Once you have saved the 
New Calibration, Field Calibration will appear 
the next time you enter Calibration Mode.

Factory Calibration (FACT) 

Enter Factory Calibration to reset the unit to 
its most recent factory calibration parameters. 
  1.  Press the Menu/Enter button           and 

scroll using the up or down arrow buttons 

G H 

until ‘RECAL’ is flashing on the dis-

play. Then press Menu/Enter           again.

  2.  Then the menu choices will be FACT/

FIELD/NEW. Press the up or down arrow 
buttons 

G H 

until ‘FACT’ is flashing on 

the display.

  3.  When ‘FACT’ is flashing on the display, 

press the Menu/Enter button           . 

  4.  Then with ‘SAVE’ flashing on the display 

press the Menu/Enter button           .

  5.  This will restore the most recent factory 

calibration.

  6.  If ‘SAVE’ is flashing and you do not wish 

to ‘SAVE’ the factory calibration, use the 
up or down arrow buttons 

G H 

to scroll 

to ‘CANCL’ and while ‘CANCL’ is flashing 
on the display, press Menu/Enter              
which will return you to the main menu. 

New Field Calibration (NEW):

Enter New Field Calibration if you wish to 
perform a field calibration of the GXLdp.
  1.  Press the Menu/Enter button           and 

scroll using the up or down arrow buttons    

G H 

until ‘RECAL’ is flashing on the 

display. Press the Menu/Enter Button   
          .

  2.  There will be three menu choices FACT/

FIELD/NEW.  Scroll using the up or down 
arrow buttons 

G H 

until ‘NEW’ is flashing.

  3.  When ‘NEW’ is flashing on the display, 

press the Menu/Enter Button           .

  4.  ‘ADJST/ZERO’ will now be flashing on the 

display. This will allow you to calibrate the 
Zero value for the GXLdp. 

  5.  Apply zero pressure to the Hi and Lo 

ports. This can be best accomplished 
by shorting the HI and LO ports on the 
SpoolCal

TM

 actuator tool using the silicon 

tubing supplied.

  6.  Using the up or down arrow buttons 

G H 

scroll until the display reads zero. Then 
press the Menu/Enter button           . The 
unit is now zeroed. 

  7.  ‘ADJST/SPAN’ will now be flashing on the 

display. 

  8.  Remove the silicone tubing from the 

SpoolCal

TM

 actuator tool.

  9   You will need an external pressure source 

to adjust the span setting.

  10.  Connect tubing from the pressure source 

to both the Hi and Lo pressure ports of 
the SpoolCal

TM

 actuator tool. 

 

11.  Using the external pressure source, 

increase the pressure to the full-scale value 
of the GXLdp. Then scroll using the up or 
down arrow buttons 

G H 

until the display 

reads the same as the pressure reference. 

Содержание GXLdp

Страница 1: ...USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2021 Ashcroft Inc I M011 10287_RevA_01 24 21 Installation and Operating Instructions...

Страница 2: ...2 2...

Страница 3: ...S Warnings General Information 4 Keypad Functions LCD Display 5 Menu Functions Programming 6 14 Error Messages 15 Display Resolution Table 16 Wiring Diagrams 17 18 Specifications 19 General Dimensions...

Страница 4: ...susceptible to damage when exposed to static electrical charges To avoid damage to the transducer the operator installer should follow proper ESD electrostatic discharge protection procedures before h...

Страница 5: ...er alphanumeric display portion reads the engineering units during normal operation SW1 is present if SW1 Switch 1 is active ON or OFF displays depending on trip status of the switch SW2 is present if...

Страница 6: ...y when the Backlight function is being turned off After a couple of seconds the BL OFF message will disappear from the screen Up Arrow Button G The up arrow G generally scrolls up in the menu tree or...

Страница 7: ...of the GXLdp RESET Reset Allows user to reset the GXLdp to its original factory settings ZERO Zero Allows the user to perform a user zero offset of the GXLdp SWITC Optional Switch function Allows the...

Страница 8: ...up or down arrow buttons G H until the desired dampening time is flashing Then press Menu Enter Then SAVE will be flashing Press Menu Enter to save the new DAMP setting The unit will then revert to l...

Страница 9: ...uttons G Hscroll until the display reads zero Then press the Menu Enter button The unit is now zeroed 7 ADJST SPAN will now be flashing on the display 8 You will need an external reference pres sure s...

Страница 10: ...menu and no settings will be changed ZERO Zeroing function In addition to being able to Zero from the live display the user can also Zero the unit through the Menu 1 Press Menu Enter and scroll using...

Страница 11: ...SW1RS will be flashing on the display Press the Menu Enter button again SW1 setup is now complete 15 Then SW2 DISAB will flash on the dis play If you also want to enable switch two SW2 repeat steps 4...

Страница 12: ...ATE KEY 90 COUNTER CLOCKWISE TO MONITOR POSITION From the OFF position the SpoolCal actuator tool can be inserted and removed Rotate the actuator tool 90 clockwise for Calibration mode CAL or 90 count...

Страница 13: ...Menu Enter button 5 This will restore the most recent factory calibration 6 If SAVE is flashing and you do not wish to SAVE the factory calibration use the up or down arrow buttons G H to scroll to C...

Страница 14: ...ing on the display 3 When FIELD is flashing on the display press the Menu Enter button 4 Then with SAVE flashing on the display press the Menu Enter button 5 This will restore the most recent field ca...

Страница 15: ...3 Wire Power Error Voltage output is selected but there is no external power This may be field correctable Err07 Range data missing or corrupt FATAL ERROR LIMIT LIMIT will display if the user tries to...

Страница 16: ...0 0 250 00 160 Pa 160 0 160 0 300 Pa 0 0 300 00 200 Pa 200 0 200 0 400 Pa 0 0 400 00 300 Pa 300 0 300 0 500 Pa 0 0 500 00 400 Pa 400 400 600 Pa 0 0 600 00 500 Pa 500 500 1 00 kPa 0 000 1 000 600 Pa 60...

Страница 17: ...Side current sense shown in dashed lines Power Vdc V V Vout EXT S1 S2 Vdc 3 Wire 4 20mA without switch Power Vdc V V Vout EXT S1 S2 Vdc 3 Wire 4 20mA Output w NPN Switches Power Vdc V V Vout EXT S1 S...

Страница 18: ...PN Load 3 Wire Voltage Output Power Vdc Vac V V V Vout EXT S1 S2 Vdc 18 3 Wire Voltage Output With NPN Switches 3 Wire Voltage Output 3 Wire Voltage Output with PNP Switches GXLdp Wiring Diagrams 3 Wi...

Страница 19: ...Vdc Vac typical Reverse polarity and mis wire protected LCD Display 4 digits for most ranges The resolution decimal place is fixed based on the range and is not field changeable See display resolutio...

Страница 20: ...20 1 34 TYP 1 59 5 49 1 95 47 50 21 5 14 1 86 2 68 2 70 1 45 R 10 R5 24 HIGH SIDE LOW SIDE OFF SPOOL CAL POSITION MONITOR SPOOL CAL POSITION ROTATE KEY CLOCKWISE ROTATE KEY COUNTER CLOCKWISE OFF SPOO...

Страница 21: ...a cutout 7 minimum vertical x 5 9 minimum horizontal in your panel or mounting location 2 Provide 4 6 32 threaded holes as shown in the drawings below 3 Place 3 spacers between the unit and the panel...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...ft Inc 250 East Main Street Stratford CT 06614 USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2021 Ashcroft Inc I M011 10287_RevA_01...

Отзывы: