  3.  When ‘FIELD’ is flashing on the display, 

press the Menu/Enter button           .

  4. Then with ‘SAVE’ flashing on the display 

press the Menu/Enter button            .

  5.  This will restore the most recent field 

calibration.

  6.  If ‘SAVE’ is flashing and you do not wish 

to ‘SAVE’ the field calibration, use the up 
or down arrow buttons 

G H 

to scroll to 

‘CANCL’ and while ‘CANCL’ is flashing on 
the display, press Menu/Enter           
which will return you to the main menu. 

RESET:

 

This will allow the user to revert to all menu 
settings as shipped from the factory.
This includes: Analog Output, Switch settings 
(if option is applicable), Damping, and User 
Zero.  This will also clear any saved Field 
Calibration from memory and reset the GXLdp 
to the Factory Calibration.

To perform the RESET procedure:
  1.  Press Menu/Enter            and scroll using 

the up or down arrows 

G H 

until ‘RESET’ 

is flashing. Then press Menu/Enter          .

  2.  ‘SAVE’ will now be flashing on the display.  

If you wish to restore all factory default 
settings, press Menu/Enter           while 
‘SAVE’ is flashing on the display. The unit 
will then return to the main menu and all 
factory settings will be re-stored including 
FACTORY Calibration.

  3.  If you do not want to restore all factory 

default settings, scroll using the up or 
down arrow keys 

G H 

until ‘CANCL’ is 

flashing on the display, and then press 
Menu/Enter          . The unit will return to 
the main menu and no settings will be 
changed.

ZERO (Zeroing function)

In addition to being able to Zero from the 
live display, the user can also Zero the unit 
through the Menu. 

  1.  Press Menu/Enter           and scroll using  

the up or down arrows 

G H 

until ‘ZERO’ 

is flashing on the display.  

  2.  Press Menu/Enter           and then while 

‘SAVE’ is flashing press  Menu/Enter  
again.  This will zero the unit, and return 
to live pressure readings. 

OPTIONAL FEATURES: 

SWITC:  Switch function

  

The GXLdp is available with an optional 
switch function.  If ordered this feature will 
allow the user to choose between 1 or 2 PNP 
or NPN switches as well as allowing the user 
to choose the set and reset point of each 
switch.  
If the user chooses to use two switches, they 
must be the same type, either PNP or NPN.

NOTE

:

To set an Increasing set point, the reset value 
must be lower than the set point
To set a Decreasing set point, the reset value 
must be higher than the set point

To enable and set the switch feature

  1.  To enable the optional switch, press the 

Menu/Enter button           , and scroll 
through the main menu using the up or 
down arrow buttons 

G H 

until ‘SWITC’ is 

flashing. Then press Menu/Enter button  
   again. 

  2.  Initially, PNP will be flashing on the dis-

play, and you can change the switch type 
to NPN by pressing the up or down arrow 
buttons 

G H

  3.  Once you are on the desired switch type 

press the Menu/Enter button            .

  4.  Initially, SW1 DISAB will be flashing on 

the display. Press the up or down arrow 
buttons 

G H 

until SW1 ENABL is flashing 

on the display. Then press the Menu/
Enter button            .

Содержание GXLdp

Страница 1: ...USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2021 Ashcroft Inc I M011 10287_RevA_01 24 21 Installation and Operating Instructions...

Страница 2: ...2 2...

Страница 3: ...S Warnings General Information 4 Keypad Functions LCD Display 5 Menu Functions Programming 6 14 Error Messages 15 Display Resolution Table 16 Wiring Diagrams 17 18 Specifications 19 General Dimensions...

Страница 4: ...susceptible to damage when exposed to static electrical charges To avoid damage to the transducer the operator installer should follow proper ESD electrostatic discharge protection procedures before h...

Страница 5: ...er alphanumeric display portion reads the engineering units during normal operation SW1 is present if SW1 Switch 1 is active ON or OFF displays depending on trip status of the switch SW2 is present if...

Страница 6: ...y when the Backlight function is being turned off After a couple of seconds the BL OFF message will disappear from the screen Up Arrow Button G The up arrow G generally scrolls up in the menu tree or...

Страница 7: ...of the GXLdp RESET Reset Allows user to reset the GXLdp to its original factory settings ZERO Zero Allows the user to perform a user zero offset of the GXLdp SWITC Optional Switch function Allows the...

Страница 8: ...up or down arrow buttons G H until the desired dampening time is flashing Then press Menu Enter Then SAVE will be flashing Press Menu Enter to save the new DAMP setting The unit will then revert to l...

Страница 9: ...uttons G Hscroll until the display reads zero Then press the Menu Enter button The unit is now zeroed 7 ADJST SPAN will now be flashing on the display 8 You will need an external reference pres sure s...

Страница 10: ...menu and no settings will be changed ZERO Zeroing function In addition to being able to Zero from the live display the user can also Zero the unit through the Menu 1 Press Menu Enter and scroll using...

Страница 11: ...SW1RS will be flashing on the display Press the Menu Enter button again SW1 setup is now complete 15 Then SW2 DISAB will flash on the dis play If you also want to enable switch two SW2 repeat steps 4...

Страница 12: ...ATE KEY 90 COUNTER CLOCKWISE TO MONITOR POSITION From the OFF position the SpoolCal actuator tool can be inserted and removed Rotate the actuator tool 90 clockwise for Calibration mode CAL or 90 count...

Страница 13: ...Menu Enter button 5 This will restore the most recent factory calibration 6 If SAVE is flashing and you do not wish to SAVE the factory calibration use the up or down arrow buttons G H to scroll to C...

Страница 14: ...ing on the display 3 When FIELD is flashing on the display press the Menu Enter button 4 Then with SAVE flashing on the display press the Menu Enter button 5 This will restore the most recent field ca...

Страница 15: ...3 Wire Power Error Voltage output is selected but there is no external power This may be field correctable Err07 Range data missing or corrupt FATAL ERROR LIMIT LIMIT will display if the user tries to...

Страница 16: ...0 0 250 00 160 Pa 160 0 160 0 300 Pa 0 0 300 00 200 Pa 200 0 200 0 400 Pa 0 0 400 00 300 Pa 300 0 300 0 500 Pa 0 0 500 00 400 Pa 400 400 600 Pa 0 0 600 00 500 Pa 500 500 1 00 kPa 0 000 1 000 600 Pa 60...

Страница 17: ...Side current sense shown in dashed lines Power Vdc V V Vout EXT S1 S2 Vdc 3 Wire 4 20mA without switch Power Vdc V V Vout EXT S1 S2 Vdc 3 Wire 4 20mA Output w NPN Switches Power Vdc V V Vout EXT S1 S...

Страница 18: ...PN Load 3 Wire Voltage Output Power Vdc Vac V V V Vout EXT S1 S2 Vdc 18 3 Wire Voltage Output With NPN Switches 3 Wire Voltage Output 3 Wire Voltage Output with PNP Switches GXLdp Wiring Diagrams 3 Wi...

Страница 19: ...Vdc Vac typical Reverse polarity and mis wire protected LCD Display 4 digits for most ranges The resolution decimal place is fixed based on the range and is not field changeable See display resolutio...

Страница 20: ...20 1 34 TYP 1 59 5 49 1 95 47 50 21 5 14 1 86 2 68 2 70 1 45 R 10 R5 24 HIGH SIDE LOW SIDE OFF SPOOL CAL POSITION MONITOR SPOOL CAL POSITION ROTATE KEY CLOCKWISE ROTATE KEY COUNTER CLOCKWISE OFF SPOO...

Страница 21: ...a cutout 7 minimum vertical x 5 9 minimum horizontal in your panel or mounting location 2 Provide 4 6 32 threaded holes as shown in the drawings below 3 Place 3 spacers between the unit and the panel...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...ft Inc 250 East Main Street Stratford CT 06614 USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2021 Ashcroft Inc I M011 10287_RevA_01...

Отзывы: