15
GXLdp Error Messages
Note: Multiple Errors may be displayed.
Err02
Configuration Error. Mismatch between sensor serial number
on the sensor board and the sensor serial number recorded
on the display board. FATAL ERROR.
Err03
Calibration Error. Device has not been factory calibrated or the
factory calibration data is corrupt. FATAL ERROR.
Err04
Settings Error. User Settings (Menu settings) are corrupt.
FATAL ERROR.
Err05
Communication Error. Pressure sensor has stopped working.
FATAL ERROR.
Err06
3-Wire Power Error. Voltage output is selected, but there is no
external power. This may be field correctable.
Err07
Range data missing or corrupt. FATAL ERROR
LIMIT
LIMIT will display if the user tries to Zero more than 5% of
Span. LIMIT will also display if the user tries to adjust Span by
more than 5% of the positive full-scale value.
OVER
OVER will display if the unit is Over-Pressured.
S LOC
S LOC will display if the user attempts to make a span adjust-
ment at less than 90% of the positive full-scale value
.
UNDER
UNDER will display if the unit is Under-Pressured.
Z LOC
Z LOC will display if the user attempts to make a zero offset
adjustment while outside of Zero ±10% of positive full-scale.