background image

11

  5.  SW1SP will be flashing on the display.  

This will allow the user to calibrate the 
SW1 (Switch 1) set point.

  6.  Press the Up arrow 

G

, this will cause the 

first digit to flash.

  7.  If you wish to change this digit press the 

up or down arrow buttons 

G H  

(NOTE: for Bidirectional ranges you can 
set the switch for a negative set point, 
this will be designated by a minus sign 
showing in front of the set point on the 
display).

  8.  Once you adjust the digit to the appropri-

ate number, press the Menu/Enter button      
          to get to the next digit.

  9.  Repeat steps 7 & 8 until all 5 digits are 

displaying correctly. While the last digit is 
flashing, press the Menu/Enter button 
          . Then SW1SP will be flashing on 
the display. Press the Menu/Enter button 
          again.  

  10.  SW1RS will be flashing on the display.  

This will allow the user to calibrate the 
Switch 1 reset point.

  11.  Press the Up arrow 

G

, this will cause 

the first digit to flash.

  12.  To change this digit press the up or 

down arrow buttons 

G H  

(NOTE: for Bidirectional ranges you can 
set the switch for a negative set point, 
this will be designated by a minus sign 
showing in front of the set point on the 
display).

  13.  Once you adjust the digit to the appro-

priate number, press the Menu/Enter 
button           to get to the next digit.

  14.  Repeat steps 12 & 13 until all 5 digits 

are displaying correctly.  While the last 
digit is flashing, press the Menu/Enter 
button           . Then SW1RS will be 
flashing on the display.  Press the Menu/
Enter button           again. SW1 setup is 
now complete.

  15.  Then SW2 DISAB will flash on the dis-

play.  If you also want to enable switch 

two (SW2) repeat steps 4 through 14 to 
setup SW2. If you do not wish to enable 
SW2, simply press the Menu/Enter but-
ton           while SW2 DISAB is flashing.

 

16.  SAVE will flash on the display. Press 

the Menu/Enter button           to save 
the settings for both SW1 and SW2. The 
display will return to the main menu.

NOTE:

 Once you have completed the set up 

for the SWITCH (SWITC) function, any enabled 
switch is indicated in the lower left portion of 
the display. If SW1 or SW2 are energized the 
display will show ON under the appropriate 
switch, if SW1 or SW2 are de-energized the 
display will show OFF under the appropriate 
switch.

SpoolCal

TM

 (XPV):

The rotating process valve actuator in 
conjunction with the SpoolCal™ actuator tool 
provides two types of available tests including 
Calibrate (CAL) and Monitor (MONITOR) 
through the appropriate Hi and Lo Port of the 
SpoolCal actuator tool. In the CAL (Cali-
brate) mode, the GXLdp is isolated from the 
process and allows externally generated test 
pressure input for calibration. In the MONITOR 
(Monitor) mode, the live system pressure can 
be monitored with an ASHCROFT ATE-2 Cal-
ibrator or other handheld instrument without 
physically unplugging the process tubes.  
The “XPV” option supplies a SpoolCal™ 
actuator and 7˝ of silicon tubing. The actuator 
tool can be set to either CAL (Calibrate mode), 
MONITOR (Monitor mode) or OFF, and also 
includes HI (high) and LO (low) differential 
pressure ports.

Содержание GXLdp

Страница 1: ...USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2021 Ashcroft Inc I M011 10287_RevA_01 24 21 Installation and Operating Instructions...

Страница 2: ...2 2...

Страница 3: ...S Warnings General Information 4 Keypad Functions LCD Display 5 Menu Functions Programming 6 14 Error Messages 15 Display Resolution Table 16 Wiring Diagrams 17 18 Specifications 19 General Dimensions...

Страница 4: ...susceptible to damage when exposed to static electrical charges To avoid damage to the transducer the operator installer should follow proper ESD electrostatic discharge protection procedures before h...

Страница 5: ...er alphanumeric display portion reads the engineering units during normal operation SW1 is present if SW1 Switch 1 is active ON or OFF displays depending on trip status of the switch SW2 is present if...

Страница 6: ...y when the Backlight function is being turned off After a couple of seconds the BL OFF message will disappear from the screen Up Arrow Button G The up arrow G generally scrolls up in the menu tree or...

Страница 7: ...of the GXLdp RESET Reset Allows user to reset the GXLdp to its original factory settings ZERO Zero Allows the user to perform a user zero offset of the GXLdp SWITC Optional Switch function Allows the...

Страница 8: ...up or down arrow buttons G H until the desired dampening time is flashing Then press Menu Enter Then SAVE will be flashing Press Menu Enter to save the new DAMP setting The unit will then revert to l...

Страница 9: ...uttons G Hscroll until the display reads zero Then press the Menu Enter button The unit is now zeroed 7 ADJST SPAN will now be flashing on the display 8 You will need an external reference pres sure s...

Страница 10: ...menu and no settings will be changed ZERO Zeroing function In addition to being able to Zero from the live display the user can also Zero the unit through the Menu 1 Press Menu Enter and scroll using...

Страница 11: ...SW1RS will be flashing on the display Press the Menu Enter button again SW1 setup is now complete 15 Then SW2 DISAB will flash on the dis play If you also want to enable switch two SW2 repeat steps 4...

Страница 12: ...ATE KEY 90 COUNTER CLOCKWISE TO MONITOR POSITION From the OFF position the SpoolCal actuator tool can be inserted and removed Rotate the actuator tool 90 clockwise for Calibration mode CAL or 90 count...

Страница 13: ...Menu Enter button 5 This will restore the most recent factory calibration 6 If SAVE is flashing and you do not wish to SAVE the factory calibration use the up or down arrow buttons G H to scroll to C...

Страница 14: ...ing on the display 3 When FIELD is flashing on the display press the Menu Enter button 4 Then with SAVE flashing on the display press the Menu Enter button 5 This will restore the most recent field ca...

Страница 15: ...3 Wire Power Error Voltage output is selected but there is no external power This may be field correctable Err07 Range data missing or corrupt FATAL ERROR LIMIT LIMIT will display if the user tries to...

Страница 16: ...0 0 250 00 160 Pa 160 0 160 0 300 Pa 0 0 300 00 200 Pa 200 0 200 0 400 Pa 0 0 400 00 300 Pa 300 0 300 0 500 Pa 0 0 500 00 400 Pa 400 400 600 Pa 0 0 600 00 500 Pa 500 500 1 00 kPa 0 000 1 000 600 Pa 60...

Страница 17: ...Side current sense shown in dashed lines Power Vdc V V Vout EXT S1 S2 Vdc 3 Wire 4 20mA without switch Power Vdc V V Vout EXT S1 S2 Vdc 3 Wire 4 20mA Output w NPN Switches Power Vdc V V Vout EXT S1 S...

Страница 18: ...PN Load 3 Wire Voltage Output Power Vdc Vac V V V Vout EXT S1 S2 Vdc 18 3 Wire Voltage Output With NPN Switches 3 Wire Voltage Output 3 Wire Voltage Output with PNP Switches GXLdp Wiring Diagrams 3 Wi...

Страница 19: ...Vdc Vac typical Reverse polarity and mis wire protected LCD Display 4 digits for most ranges The resolution decimal place is fixed based on the range and is not field changeable See display resolutio...

Страница 20: ...20 1 34 TYP 1 59 5 49 1 95 47 50 21 5 14 1 86 2 68 2 70 1 45 R 10 R5 24 HIGH SIDE LOW SIDE OFF SPOOL CAL POSITION MONITOR SPOOL CAL POSITION ROTATE KEY CLOCKWISE ROTATE KEY COUNTER CLOCKWISE OFF SPOO...

Страница 21: ...a cutout 7 minimum vertical x 5 9 minimum horizontal in your panel or mounting location 2 Provide 4 6 32 threaded holes as shown in the drawings below 3 Place 3 spacers between the unit and the panel...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...ft Inc 250 East Main Street Stratford CT 06614 USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2021 Ashcroft Inc I M011 10287_RevA_01...

Отзывы: