E-Command Table
** NOTE: Settings will be saved in unit internal eprom/memory until
changed or EE Jumper removed
The structure of the command is
EXXXX<CR><LF> w
here XXXX is the
mode setting value
E00XY
: Flow Mode.
E10XY
: Normal Foot Switch Mode, Full Comma Delimited.
E11XY
: Min/Max/TIR (PORT 1)
E12XY
: Min/Max/TIR (PORTS 1, 2)
E13XY
: Min/Max/TIR (PORTS 1, 2, 3)
E14XY
: Min/Max/TIR (PORTS 1, 2, 3, 4)
E15XY
: Min/Max/TIR (PORT 1, 2, 3, 4, 5)
E16XY
: Min/Max/TIR (PORTS 1, 2, 3, 4, 5, 6)
E17XY
: Min/Max/TIR (PORTS 1, 2, 3, 4, 5, 6, 7)
E18XY
: Min/Max/TIR (PORTS 1, 2, 3, 4, 5, 6, 7, 8)
E20XY
: Normal Foot Switch Mode, Reading Only.
E30XY
: Normal Foot Switch Mode, Mitutoyo Mux-10.
E40XY
: N/A
E50XY
: Global Foot Switch Mode, Full Comma Delimited.
E60XY
: Global Foot Switch Mode, Reading Only.
E70XY
: Global Foot Switch Mode, Mitutoyo Mux-10.
X
: Baud Rate
Y
: Data Output
0: 4800 0: USB
1: 9600 1: RS-232
2: 19200 2: Keyboard mode - 1
3: 38400 3: Keyboard mode - 2
4: Keyboard mode – 3
5: Keyboard mode – 4
6: Keyboard mode – 5
7: Keyboard mode – 6