PST & PSS & PSH SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
45
Queues
The output queue is included to power supplies.
Output Queue
: The programmable power supplies store query responses
in the output queue by succeeding the IEEE 488.2 protocol. If the power
supply receives a new command or query message after a message
terminator, the power supply will clear and reset this queue each time.
The computer must read a query response before it sends the next
command (or query) or it loses response to earlier queries.
Error/Event Q ueues
When an error or event occurs, the output queue stores the message. The
output queue stores and reports the messages on a FIFO (first in first out)
state. The
SYSTem:ERRor?
query reads the next item from the output
queue. If output queue overflows, the error message is –350,
“Queue
overflow”
; the queue can’t store or report succeeding messages till it
is read or cleared.
Error Message
Table 8 lists the SCPI error messages for the programmable power
supplies.
Table 8 T he error messages for the power supplies
SCPI Error Code and Description
SESR Bit
0, “No error”
-100, “Command error”
5
-200, “Execution Error”
4
-221, “Settings conflict ”
4
-221, “Settings conflict ; Timer setting error”
4
-221, “Settings conflict ; Overvoltage protection setting error”
4
-221, “Settings conflict ; Voltage setting error”
4
PST & PSS & PSH SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
46
-221, “Settings conflict; Current setting error”
4
-221, “Settings conflict; Recall setting error”
4
-221, “Settings conflict; Store setting error”
4
-222, “Data out of range ”
4
-222, “Data out of range ; Voltage too large”
4
-222, “Data out of range ; Current too large”
4
-222, “Data out of range ; Voltage too small”
4
-222, “Data out of range ; Current too small”
4
-240, “Hardware Error”
4
-300, “Device-specific error ”
3
-300, “Device-specific error; Overcurrent protection error”
3
-300, “Device-specific error; Overvoltage protection error ”
3
-300, “Device-specific error; Overtemperature protection
error”
3
-300, “Device-specific error; Calibration current error”
3
-300, “Device-specific e rror; Calibration voltage error ”
3
-300, “Device-specific error; Calibration overvoltage
protection error ”
3
-310, “System error”
3
-313, “Calibration memory lost”
3
-330, “Self-test failed ”
3
-330, “Self-test failed ; CPU test error”
3
-330, “Self-t est failed ; RAM test error”
3
-330, “Self-test failed ; ROM test error”
3
-330, “Self-test failed ; DAC/ADC test error”
3
-350, “Queue overflow ”
3
-410, “Query INTERRUPTED ”
2
-420, “Query UNTERMINATED ”
2
-430, “Query DEADLOCKED ”
2