Model 8640B
TM 9-4935-601-14-7&P
PERFORMANCE TESTS
4-11. FREQUENCY RANGE TEST
SPECIFICATION:
Range: 500 kHz to 512 MHz in 10 octave bands.
Bands and Band Overlap: Bands extend 10% below and 7% above the nominal limits shown below.
Nominal
Frequency
External
Bands
0.5-1
1-2
2-4
4-8
8-16
16-32
32-64
64-
128-
256-
Doubler Band
(MHz)
128
256
512
512-1024
Frequency
0.45
0.9
1.8
3.6
7.2
14.4
28.8
57.5
115
230
230 to 550
Range (MHz)
to
to
to
to
to
to
to
to
to
to
(without Ex-
(with overlap)
1.07
2.1
4.2
8.5
17.1
34.3
68.7
137.5
275
550
ternal Doubler)
DESCRIPTION:
The frequency range is verified by using a frequency counter to measure the frequency at the high and low end of each
band.
EQUIPMENT:
Frequency Counter .....................................................................................HP 5327C
PROCEDURE:
1.
Connect generator's auxiliary RF output jack (located on rear panel) to frequency counter's 50 ohm input after
setting Signal Generator's controls as follows:
COUNTER MODE:
EXPAND ................................................................................... Off
LOCK........................................................................................ Off
Source ..................................................................................... INT
AM............................................................................................................................................................ OFF
FM............................................................................................................................................................ OFF
RANGE ...................................................................................................................................... 256-512 MHz
FREQUENCY TUNE ................................................................................................................. Full clockwise
FINE TUNE .......................................................................................................................................Centered
RF ON/OFF................................................................................................................................................ON
2.
Set FREQUENCY TUNE full clockwise. The frequency counter should read 550 MHz or greater.
550.0 MHz _______________________
4-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com