Model 8640B
TM 9-4935-601-14-7&P
PERFORMANCE TESTS
4-35. PEAK INCIDENTAL PHASE MODLULATION TEST
SPECIFICATION:
Peak Incidental PM (at 30% AM):
Less than 0.15 radians, 0.5 to 128 MHz.
Less than 0.3 radians, 128 to 512 MHz.
DESCRIPTION:
A vector voltmeter is used to compare the phase of the signal into the generator's modulation amplifier with the phase of
the same signal (modulated at a 0.1 Hz rate) at the RF OUTPUT port. The signal is supplied by the generator's own
oscillator and divider circuits and has low incidental PM.
Figure 4-21. Peak Incidental Phase Modulation Test Setup
EQUIPMENT:
Function Generator ....................................................................................
HP 3300A
Vector Voltmeter ........................................................................................
HP 8405A
50 Ohm Tee (2 required)............................................................................
HP 11536A
50 Ohm Load ............................................................................................
HP 11593A
Test Cable (2 required) ....................................................................
HP 11592-60001
Adapter ...............................................................................................
HP 1250-0827
4-54