5300 Series
TECSource
User’s Manual · Page 39
Error
Code
Description
Cause
E-415 Sensor
short,
output
turned off
A sensor short circuit was detected and the
output was turned off.
E-416 Cal failed
An error occurred during calibration that
caused the calibration to fail. The unit is no
longer in calibration mode.
E-419 TEC not stable
The TEC is considered stable if the
temperature has changed less than 0.02°C
for more than 20 seconds.
E-433 Not a TEC
The TEC:CHAN command attempted to
select a non-TEC channel
E-434 Ite limit exceeds
cable rating
The cable plugged into the unit is not rated
to carry the current as set by the current
limit. Use a 1262B cable, 1263B cable, or
custom cable with ID pin connected to
analog ground, or reduce the current limit
below 5A.
E-435 Mode Change
A mode change occurred when the output
was on, forcing the output off.
E-436 AutoTune failed
The AutoTune process failed. See the
“Using the AutoTune Function” for more
information.
E-437 AutoTune in T mode
only
The AutoTune process is only available in T
(temperature) mode. In R mode, the PID
parameters much be determined manually.
E-997 Control error, cycle
power
A hardware control error occurred, cycle
power to resolve. If error continues to occur,
contact factory.
E-998 Command
not
supported
A command was received that is not
supported by this instrument.
E-999 Non-specific error
A non-specific error was encountered.
Содержание TECSource 5305
Страница 1: ......
Страница 35: ...5300 Series TECSource User s Manual Page 35...
Страница 43: ...5300 Series TECSource User s Manual Page 43 NOTES...