5300 Series
TECSource
User’s Manual · Page 13
Settings and Menus
All parameters of the
TECSource
can be viewed and changed within the menu.
The menus are constructed with the most used parameters first. To change any
setting, press
Menu
to enter the menu then rotate the knob to select the
parameter to change. Press the knob to begin changing the value. As a visual
indication that you are in edit mode, you will see an asterisk appear next to the
value. Once you have made your change, press the knob or
Menu
button to
store the value. Pressing the
Menu
button will store and exit the menu, while
pressing the knob will store the value but leave you in the menu to make
additional changes.
Some settings are contained inside a sub menu, such as communications
settings. To access the sub menu, simply press the knob to enter the sub menu
when its name is displayed.
Below is a complete list of available settings:
Menu
Description
Factory
Default
Root Menu
Main Menu
Mode
This set the operating mode (
T Mode
,
R Mode
,
or
Ite Mode
) of the temperature controller.
T Mode
Mount
Specify the mount connected to the
TECSource
. If using a
LaserMount
, select the
appropriate model. Otherwise, select
User
Defined
.
User Defined
I Lim
I Lim
sets the current limit of the temperature
controller. The limit should be set to a value
that is suitable for your Peltier device.
3A
Gain
Gain
controls the response of the temperature
controller. A higher gain value will cause the
controller to respond more quickly to the
difference between the set point and the actual
temperature, while a lower value will cause it to
respond more slowly. Select
PID
for direct
access to the PID parameters. Read more on
setting gain below.
30
PID P
The proportional term of the PID loop. Will only
be available if
Gain
is set to
PID
.
1