Page 36 · 5300 Series
TECSource
User’s Manual
Specifications
All values relate to a one-hour warm-up period.
Description
5305
5310
5300-08-24
Drive Channel
Current
Range (A)
5
10
8
Compliance Voltage (V)
12
12
24
Max Power (W)
60
120
192
Resolution (A)
0.01
0.01
0.01
Accuracy (
±
[% set point + mA])
0 + 30
0 + 60
0 + 60
Noise/Ripple (mA, rms)
< 5
< 15
< 15
Temperature Control
Range (ºC)
1
-99
to
250
Resolution (ºC)
0.01
Thermistor Accuracy (± ºC)
2
0.05
3
AD560 Accuracy (± ºC)
2
0.05
LM335 Accuracy (± ºC)
2
0.05
RTD Accuracy (± ºC)
2
0.05
Short Term Stability (1hr) (± ºC)
4
0.004
Short Term Stability (24hr) (± ºC)
4
0.01
Measurement Channels
Current
Resolution (mA)
10
Accuracy (
±
[% r mA])
0 + 30
0 + 60
0 + 60
Voltage
Resolution (mV)
10
Accuracy (
±
[% r V])
0 + 0.05
Sensor
10
μ
A Thermistor
Range (k
Ω
)
0.1 – 450
Resolution (k
Ω
) 0.01
Accuracy (
±
[% r k
Ω
]) 0.05
+
50
100
μ
A Thermistor
Range (k
Ω
) 0.05
–
45
Resolution (k
Ω
) 0.001
Accuracy (
±
[% r k
Ω
]) 0.05
+
5
LM335
Bias (mA)
1
Range (mV)
1730 – 4730
Resolution (mV)
0.1
Accuracy (
±
[% r mV])
0.3 + 1
1
Software limits. Actual range dependent on sensor type and system dynamics
2
Accuracy figures are the additional error the 5300 adds to the measurement, and does not include the
sensor uncertainties.
3
25°C, 100
μ
A thermistor
4
Stability measurements done at 25°C using a 10k
Ω
thermistor on the 100
μ
A setting. The number is ½
the peak-to-peak deviation from the average over the measurement period.