![Armedica AM-700 Series Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/armedica/am-700-series/am-700-series_manual_2973752002.webp)
AM-700 SERIES PARALLEL BARS
ELECTRIC PARALLEL BARS INSTALLATION
AM-700
1. Remove platform (1) from crate using handles attached to mounting brackets. (See item
#66 on AM700 installation diagram). 4 people are required. (9). Set the platform on
supports to get it off the floor far enough to allow access to the bottom of the platform
while installing control stand.
DO NOT LIFT OR SUPPORT BY END RAMPS OR COLUMNS!!
2. Place the four upright columns (4) between the brackets on the platform according to letter
designations (9) and tighten the mounting bolts (8) securely.
3. Install the hand rails (2) onto the adjustment tubes (14) using spacer (5), washers (6), and
lock nut (7); tighten the lock nut (7), sufficiently, to eliminate any play but not so tight that
they will not pivot when adjusting the width. When necessary, adjust the width according to
the operating instructions.
4. Remove the cover plate from box (26) to allow access to inside of control stand (11).
Using washers (44), and nuts (13), mount the control stand (11) to the platform (1),
threading controller power cable, and connector cables (16, 17) through appropriate holes.
5. With screws (12) and cover plate (26) removed, connect the controller (3) per the following
instructions.
DO NOT CONNECT THE CONTROL BOX TO POWER SOURCE BEFORE READING THESE
INSTRUCTIONS OR DAMAGE MAY OCCUR!!!
Connect the four cables (16) and (17) (DIN8 connectors) to the appropriate column (4).
Connect the four cables (16) and (17) (DIN8 connectors) to the controller (3).
The cables must be fully inserted so the arrows are completely inside the control
box receptacle or the legs will malfunction and may cause permanent damage.
Verify that the control switch (10) cable with a blue DIN7 socket marked with a blue “HS” is
properly connected to the controller (3).
Verify that the required supply voltage indicated on the type plate corresponds with the
supply voltage of the electrical outlet.
Connect the unit to the electrical outlet.