• The GLUCOCARD Vital Blood Glucose Monitoring System is not designed to
substitute for pathology laboratory equipment and should not be used for
the diagnosis of diabetes.
• Severe dehydration (excessive water loss) may cause false low results. If
you believe you are suffering from dehydration, consult your healthcare
professional immediately.
• A red blood cell count (hematocrit) that is very high (above 52%) can give
abnormally low results. A very low hematocrit level (below 33%) can give
abnormally high results.
• Inaccurate results may occur in severely hypotensive individuals or patients
in shock.
• Inaccurate low results may occur for individuals experiencing a
hyperglycemic-hyperosmolar state, with or without ketosis.
• Patients undergoing oxygen therapy may yield false results.
• Do not use to test newborns. The system has not been validated for
neonatal use.
WARNINGS, PRECAUTIONS, AND LIMITATIONS