Thank you for choosing GLUCOCARD

TM

 SM. 

We have designed this compact blood glucose meter for faster and easier 
testing of your blood glucose. We hope our product will help you manage your 
diabetes. This manual explains how to use your new meter. Before testing, 
carefully read through this manual and the package inserts that come with the 
GLUCOCARD

TM

 SM TEST STRIPS and GLUCOCARD

TM

 SM CONTROL. Pay 

particular attention to listed warnings and cautions. Please keep this manual 
handy for future reference. If you have any questions, please contact  

 

A. Menarini Diagnostics on the numbers listed at the end of the manual.

Chapter 1  Introduction

1.1 Intended use

The GLUCOCARD

TM

 SM meter and GLUCOCARD

TM

 SM TEST STRIPS are for 

quantitatively measuring the glucose level in fresh capillary whole blood. They 
are intended for use outside of the body (

in vitro

 diagnostic use) at home or in a 

clinical setting as an aid to monitor and control blood glucose levels. 

Do not use them to diagnose diabetes

. Also, 

do not

 alter treatment based 

on the test results of this meter without instructions from your doctor or 
healthcare professional.

1.2 Measurement principle

Glucose in the blood reacts with the reagent in the test strip and this produces 
a small electric current. The strength of this current is proportional to the 
concentration of glucose in the blood. The meter measures this current and 
calculates your blood glucose level.

1.3 What’s included

„

 GLUCOCARD

TM

 SM

„

 Carrying case

„

  Operating manual (this manual)

„

  Guide on Bluetooth connection

„

 GLUCOCARD

TM

 SM TEST STRIPS

„

 Lancets

„

 Lancing device

Chapter 2  Before tests

2.1 Important health-related information

 WARNING

Do not

 use this meter for your blood glucose testing if you are being tested 

for xylose absorption.

These treatments may produce higher test results than your actual 
blood glucose level

.

2.2 What you need for each test

a. GLUCOCARD

TM

 SM

(meter)

b. GLUCOCARD

TM

 SM 

TEST STRIPS

c. Lancing device

d. Lancet

 WARNING

Keep the meter, test strips and other items away from young children. Small 
items are choking hazards.

Operating manual

2.3 Cautions on using the meter

 

CAUTION

„

  For accurate test results, allow the meter to adjust to the surroundings:

    Temperature: 8 to 40 °C (46 to 104 °F)
    Humidity: 20 to 80% RH (Relative Humidity)
  for 30 minutes before testing your blood glucose. 

„

  Do not store or use the meter where: 

  
  • Humidity is high enough to cause condensation (bathrooms, drying 

rooms, kitchen, etc.).

 • 

 

„

  Do not use the meter after it has been dropped in liquid or liquids have 

entered inside, even if dried afterwards.

„

  Avoid hand contact with test strip slot on the meter. A thermo sensor is 

housed inside the meter to minimize errors.

„

  Do not connect the meter to a computer via USB cable during testing. The 

meter may be damaged, leading to inaccurate test results.

„

  Do not apply blood directly to the test strip slot on the meter.

„

  Do not share your meter with anyone to avoid the risk of infection.

2.4 Cautions on using the test strips

 

CAUTION

„

  Use only the GLUCOCARD

TM

 SM TEST STRIPS for testing with the 

GLUCOCARD

TM

 SM meter. 

Do not

 use other test strips as it causes 

inaccurate test results. 

„

 

Do not

 use test strips beyond their expiration date. The expiration date is 

written on the test strip bottle next to the “Use by” symbol. 

„

  For accurate test results, allow test strips to adjust to the surroundings:

    Temperature: 8 to 40 °C (46 to 104 °F)
    Humidity: 20 to 80% RH (Relative Humidity)
  for 30 minutes before testing your blood glucose.  

„

 

Do not 

use the test strips if 6 months or more have passed since opening 

the laminated pouch. 

„

  The test strips are for single use only. 

Do not

 use test strips that have 

already absorbed blood or control solution.

NOTE

Store test strips in their original bottle to maintain quality. 

Do not

 transfer them 

to other containers.

2.5 Meter parts

Display

Test strip 
ejector

Battery cover

Test strip slot light

Micro-B USB 
port

Test strip slot

Insert a test strip here 
before testing.

      

button

(Down button)

      

button

 

(ON/OFF button)

      

button

(Up button)

2.6 Display

At each start-up, the screen displays all segments, indicating the correct 
functioning of the meter.

NOTE

After 30 seconds of inactivity, the display will dim slightly. It will brighten again 
when next used.

Result average symbol

High blood sugar (high 

value) warning flag

Alarm symbol

Hospital device icon

Date symbol

Time symbol

Delete flag

Past results symbol

Low blood sugar (low

value) warning flag

Battery symbol

Beeper volume symbol

Control test flag

Sound mute symbol

Before bedtime flag

Blood symbol:

 

Appears when the meter 
is ready for testing. 

Bluetooth symbol:

Flashes during wireless 
communication 

Personal device icon

Before-meal flag:

indicates the “After-meal flag”.

Thermometer symbol

Setting symbol:

Appears when you set up the meter.

Wireless communication setting symbol

Appears when you set up the wireless 
communication (BLE = Bluetooth Low Energy).

Memory symbol:

 

Appears when you are 
reviewing a past result 
from the meter memory.  

BLOOD GLUCOSE MONITORING SYSTEM

Содержание GLUCOCARD SM

Страница 1: ...e condensation bathrooms drying rooms kitchen etc Do not use the meter after it has been dropped in liquid or liquids have entered inside even if dried afterwards Avoid hand contact with test strip slot on the meter A thermo sensor is housed inside the meter to minimize errors Do not connect the meter to a computer via USB cable during testing The meter may be damaged leading to inaccurate test re...

Страница 2: ...lucose If nothing appears on the screen remove the test strip and insert it back into 2 Touch the tip of the test strip to the drop of blood with blood Accurate test results will not be obtained if you apply your blood to the check window directly The meter counts down After 5 seconds a beep sounds and the test result and date time are shown NOTE For accurate test results touch the test strip to y...

Страница 3: ...ph 5 2 2 Press 9 JQ 3 9 JQ button The result average changes in the order of 7 14 30 60 and 90 day each time button is pressed Three bars appear when there is no past result to be included in the average calculation 4 button 5 To turn off the meter hold down button for 3 seconds 5 4 Deleting all test results You can delete all test results from the memory Remember that results cannot be retrieved ...

Страница 4: ...ow the threshold value 10 to 120 mg dL HYPO Current setting HYPER To the top of the HYPER setting 1 Turn on the HYPER flag 2 Set the threshold value A HYPER flag appears when your test results are above the threshold value 150 to 600 mg dL Beeper volume On1 Low On2 Normal On3 High OFF Off 1 Select a setting To the top of the beeper volume setting Date 1 Set the year To the top of the date setting ...

Страница 5: ...rocedure 2 Select PrG Selection of a different receiving device 1 Select the number of the receiving device 3 Edit the icon assigned optional 2 Select On To the BLE mode selection SUCCESS FAILURE Back to step 3 Check the settings of your receiving device Turning the wireless communication OFF 1 Select OFF To the BLE mode selection Deletion of a receiving device 1 Select the number of the receiving...

Страница 6: ...lf off Chapter 9 System care 9 1 Storing your system After use tightly close the caps of the test strip bottle and control solution bottle to maintain their quality Store your meter test strips control solutions and manuals in your carrying case and keep this in a dry location The correct storage temperatures are 0 to 50 C 32 to 122 F for the meter 1 to 30 C 34 to 86 F for test strips and 2 to 30 ...

Страница 7: ...mometer symbol and cannot be part of your result averages Chapter 11 Technical information Product GLUCOCARDTM Model SM Test item Blood glucose level Sample Fresh capillary whole blood Sample size _ Assay method dehydrogenase FAD GDH based method Mediator hexaammineruthenium III chloride 1 Methoxy PMS Calibration The results are equivalent to plasma glucose concentration capillary plasma reference...

Страница 8: ...uthorized Representative in the European Community Serial number Consult instructions for use Blood glucose meter revision Direct current voltage Catalogue number ISSUED 2015 07 Distributed by A MENARINI Diagnostics Via Sette Santi 3 50131 Firenze Italy Tel 39 XXX XXXXXXX E mail xxxxxx xxxxxxxxxx 47430 07 15 IVD Directive 98 79 EC GLUCOCARD SM Made in Philippines GLUCOCARD SM TEST STRIPS GLUCOCARD...

Отзывы: