Chapter 8  Control test

This meter checks itself every time you insert a test strip. You do not need to 
run frequent control tests. Do a control test if:

„

  You suspect the meter or test strips are not working properly.

„

  You dropped the meter.

„

  You damaged the meter.

„

  Your test results do not match how you feel.

„

  You want to check performance of the meter and test strips before a blood 

glucose test. 

NOTE

 

Use only the GLUCOCARD

TM

 SM CONTROL to test your meter and test strips.

 

CAUTION

„

 

Do not

 use control solutions beyond their expiration date. The expiration 

date is written on the control solution bottles next to the “Use by” symbol. 

„

  Leave the meter, test strips and control solution at 8 to 40 °C 

  (46 to 104 °F) and 20 to 80% RH for at least 30 minutes before running 

the control test. 

„

 

Do not

 drink control solution. It is not for human consumption. 

„

 ? $ "

1. See steps 1 and 2 in paragraph 2.7.

   

screen. 

  !

the test strip to a drop of control solution.

Accurate test results will not be obtained if you apply the control solution to the 
check window directly.

The meter counts down. After 5 seconds, a beep sounds and the test result, 
`  

NOTE 

„

  Tightly close the control solution bottle.

„

 

Do not

 touch the test strip until the test result appears. 

"   

{    
the measurement results.

 

CAUTION

+      
solutions, and contact your distributor.

5. Check that your test result is within the acceptable range.

The acceptable range is shown on the label of test strip bottles. If within range, 
the meter and the test strip are working properly. If out of range, test the control 
solution again.

 

CAUTION

If you still have test results that are out of the acceptable range, stop using 
your meter, test strips and control solutions, and contact your distributor.

6. Point the test strip down. Slide the test strip ejector to eject the test 

strip. 

The meter turns itself off.

Chapter 9  System care

9.1 Storing your system

After use, tightly close the caps of the test strip bottle and control solution bottle 
to maintain their quality. 

Store your meter, test strips, control solutions and manuals in your carrying 
case and keep this in a dry location. The correct storage temperatures are 0 to 
50 °C (32 to 122 °F) for the meter, 1 to 30 °C (34 to 86 °F) for test strips, and 2 
to 30 °C (36 to 86 °F) for control solutions. Do not freeze. Avoid heat, humidity 
and direct sunlight.

 

CAUTION

To obtain accurate test results: 

„

 

Do not

 use test strips or control solutions if their bottles are broken or 

have been left open. 

„

 

Do not

 use the test strips or control solutions beyond their expiration 

dates.

9.2 Cleaning your meter and lancing device

Your meter and lancing device do not need special cleaning. If your meter and 
lancing device get dirty, wipe them with a soft piece of cloth moistened with 
water. To disinfect these items after cleaning, wipe with a soft cloth moistened 
with 70% ethanol, 70% isopropanol or diluted household bleach (0.05% sodium 
hypochlorite solution).

9.3 Changing the batteries

When the battery symbol appears permanently on the screen, the 
batteries are getting low (approximately 100 tests can be performed). 
Before using your meter, change the batteries. Date and time settings 
will not be cancelled if you insert new batteries within 30 seconds. Past 
results remain in the memory even when the batteries are changed. 

NOTE

 

Your meter uses two 1.5V alkaline LR03/AAA batteries. This type of battery is 
available in many stores. Keep spare batteries handy at all times.

1.

 Make sure the meter is off. 

2.

|  

3.

 Remove the old batteries.

NOTE

 

Do not

 use sharp objects that may damage the meter. 

4.

 Insert two AAA alkaline batteries in the correct direction.

5.

 Close the battery cover.

NOTE

 

The meter may enter date setting mode the next time it is 
turned on. In this event, set the correct date and time. 
} ~

Dispose of old batteries according to your local environmental regulations.

Содержание GLUCOCARD SM

Страница 1: ...e condensation bathrooms drying rooms kitchen etc Do not use the meter after it has been dropped in liquid or liquids have entered inside even if dried afterwards Avoid hand contact with test strip slot on the meter A thermo sensor is housed inside the meter to minimize errors Do not connect the meter to a computer via USB cable during testing The meter may be damaged leading to inaccurate test re...

Страница 2: ...lucose If nothing appears on the screen remove the test strip and insert it back into 2 Touch the tip of the test strip to the drop of blood with blood Accurate test results will not be obtained if you apply your blood to the check window directly The meter counts down After 5 seconds a beep sounds and the test result and date time are shown NOTE For accurate test results touch the test strip to y...

Страница 3: ...ph 5 2 2 Press 9 JQ 3 9 JQ button The result average changes in the order of 7 14 30 60 and 90 day each time button is pressed Three bars appear when there is no past result to be included in the average calculation 4 button 5 To turn off the meter hold down button for 3 seconds 5 4 Deleting all test results You can delete all test results from the memory Remember that results cannot be retrieved ...

Страница 4: ...ow the threshold value 10 to 120 mg dL HYPO Current setting HYPER To the top of the HYPER setting 1 Turn on the HYPER flag 2 Set the threshold value A HYPER flag appears when your test results are above the threshold value 150 to 600 mg dL Beeper volume On1 Low On2 Normal On3 High OFF Off 1 Select a setting To the top of the beeper volume setting Date 1 Set the year To the top of the date setting ...

Страница 5: ...rocedure 2 Select PrG Selection of a different receiving device 1 Select the number of the receiving device 3 Edit the icon assigned optional 2 Select On To the BLE mode selection SUCCESS FAILURE Back to step 3 Check the settings of your receiving device Turning the wireless communication OFF 1 Select OFF To the BLE mode selection Deletion of a receiving device 1 Select the number of the receiving...

Страница 6: ...lf off Chapter 9 System care 9 1 Storing your system After use tightly close the caps of the test strip bottle and control solution bottle to maintain their quality Store your meter test strips control solutions and manuals in your carrying case and keep this in a dry location The correct storage temperatures are 0 to 50 C 32 to 122 F for the meter 1 to 30 C 34 to 86 F for test strips and 2 to 30 ...

Страница 7: ...mometer symbol and cannot be part of your result averages Chapter 11 Technical information Product GLUCOCARDTM Model SM Test item Blood glucose level Sample Fresh capillary whole blood Sample size _ Assay method dehydrogenase FAD GDH based method Mediator hexaammineruthenium III chloride 1 Methoxy PMS Calibration The results are equivalent to plasma glucose concentration capillary plasma reference...

Страница 8: ...uthorized Representative in the European Community Serial number Consult instructions for use Blood glucose meter revision Direct current voltage Catalogue number ISSUED 2015 07 Distributed by A MENARINI Diagnostics Via Sette Santi 3 50131 Firenze Italy Tel 39 XXX XXXXXXX E mail xxxxxx xxxxxxxxxx 47430 07 15 IVD Directive 98 79 EC GLUCOCARD SM Made in Philippines GLUCOCARD SM TEST STRIPS GLUCOCARD...

Отзывы: