11
SECTION
B
BLOOD GLUCOSE TESTING
OBTAINING A DROP OF BLOOD
NOTE:
• We recommend that healthcare professionals use any auto-disabling, single-use safety
lancet to obtain patient blood samples. These instructions are for the Assure
®
Lance
Safety Lancet. Call Technical Customer Service at 800.818.8877, option 5 to find a
distributor.
• We advise healthcare professionals to wear gloves during blood testing.
• Meter system must be at room temperature for testing.
Step 1
Wash patient’s hands with soap and warm water.
Dry hands thoroughly. If alcohol wipes are used, make sure the area
is dry before taking blood sample. Residual alcohol may lead to
inaccurate readings.
Step 2
Let the arm hang down for 10 to 15 seconds. (Holding the arm below
the heart and massaging the wrist, palm and finger makes it easier
to obtain a blood drop).
Step 3
Choose a site on the side of a fingertip to minimize pain.
Step 4
Grasp the outer white edge of the Assure Lance Safety Lancet.
Twist the protective cap 360° to remove.
Do not put your thumb on the top button while removing the
cap; this prevents a misfire.
Step 5
Place lancet against side of fingertip. Press button firmly.
After use, discard in sharps container.
360°