39
Electromagnetic Compatibility (EMC)
Medical electrical equipment needs special precautions regarding EMC (Electromagnetic compatibility)
and needs to be used according to the EMC information below.
Sara Flex
is intended for the use in the electromagnetic environment speci
¿
ed below. The customer or the
user of
Sara Flex
should assure that it is used in such an environment.
WARNING
Stacking or placing other electrical equipment next to this device is not recommended,
it can interfere with the equipment’s operation and safety. Portable and mobile radio-
frequency (RF) communications equipment can interfere with this equipment operation
and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take action, such as reorienting, relocating the
equipment or shielding the location.
Guidance and manufacturer’s declaration: electromagnetic emissions (EMI)
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
Sara Flex
uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Sara Flex
is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.