10
LCD Display
(See Fig. 4)
Battery Level - see
“Battery
Instructions” on page 24
.
0.0
Usage counter
Maintenance symbol
0.0
Patient’s weight - see
“Scale
(Optional)” on page 19
.
At start up it takes a little time before the display
looks as shown in
Fig. 4
. The LCD display will only
light up when a button is pressed.
Usage counter
The usage counter shows the total added amount of
time (in hours) that the lift’s lifting arm has been
raised or lowered.
The usage counter will increase in increments of 0.1
hours whenever an additional six minutes have been
added.
NOTE
The usage counter only records during
movement of the lifting arm.
Maintenance symbol
The maintenance symbol is a reminder for the yearly
maintenance requirements for the product.
The maintenance symbol is shown after a 125 hours
of usage. This is the average time a lift is used during
one year. the maintenance symbol will
À
ash (1 sec
ON and 1 sec OFF).
When the maintenance symbol appears, the unit will
still be safe to use, but the yearly maintenance should
be performed as soon as possible.
NOTE
The technician must reset the display to ‘’0.0’’
hours when the annual inspection is performed.
This is done to visually see when next
inspection is due.
Castor Brakes
The rear castors brakes are foot operated to keep
Sara
Flex
in position.
Apply the brakes
(See Fig. 5)
:
Press down on the gray castor brake lever
(A)
.
Release the brakes (See Fig. 5):
Slightly kick up the gray castor brake lever
(B)
.
0.0
kg
0.0
Fig. 4
Fig. 5
A
B