26
Cleaning and Disinfection Instructions
The following processes are recommended, but
should be adapted to comply with the local or
national guidelines (Decontamination of Medical
Devices) which may apply within the Healthcare
Facility or the country of use. If uncertain, seek
advice from the local Infection Control Specialist.
Sara Flex
should be routinely cleaned and
disinfected between patients and at regular intervals
while in use; as is good practice for all reusable
medical devices.
If there are any questions regarding cleaning
and disinfecting the equipment, or wish to order
disinfectant
À
uid, contact ArjoHuntleigh Customer
Service. See section, Parts and Accessories. Make
sure to have the Safety Data Sheets (SDS) available
for the disinfectants being used.
WARNING
To avoid eye and skin damage, always
use protective glasses and gloves. If
contact occurs, rinse with plenty of water.
If eyes or skin becomes irritated, contact
a physician. Always read the instructions
for use and material safety data sheet of
the disinfectant.
WARNING
To prevent cross-contamination, always
follow the disinfection instructions in this
Instructions for Use.
WARNING
To avoid eye or skin irritation, never
disinfect in the presence of
a patient.
Detergents/Disinfectants
• Mild neutral detergent
• Alcohol (Ethyl or Isopropyl) 70-90%
• Chlorine (10 000ppm free chlorine)
• Phenols (as per Manufacturer’s
Recommendation)
• Quaternary Ammonium (as per Manufacturer’s
Recommendation)
• Hydrogen Peroxide (<4.5%) can be used, but
may cause discoloration on the Sara Flex.
Accessories Needed for Cleaning/
Disinfection
• Protective glasses
• Protective gloves
• Spray bottle with water
• Disposable towels
• Brush