26
Disinfecting lift hygiene chair in Prelude
Always follow these 19 steps for
proper disinfection.
Procedure to be carried out at the beginning of
the day, after each resident and at the end of the
day.
1
Put on protective gloves and glasses.
2
Open the panel door and check that the dis-
infectant bottle has an ample supply of dis-
infectant and that the hose is connected to
the suction part.
(See Fig. 1)
3
Remove all plastic parts from the chair and
place them at the bottom of the
Prelude
:
•
Lift the seat insert straight upwards.
(See Fig. 2)
•
Remove the seat halves by pressing from
the outside and lift on the inside.
(See Fig. 3)
4
Remove the back rest by turning it upwards
and then pulling it straight up.
(See Fig. 4)
5
Push the lift hygiene chair carefully as far as
possible into the Prelude.
6
Check the level of the disinfectant.
7
Detach the disinfection spray handle from
the hook.
8
Open the disinfectant control as far as it will
go and set the temperature control as low as
possible.
9
Spray disinfectant in the
Prelude
while
keeping an eye on the flow meter.
10 Check the mix ratio on the flow meter so
that it is in accordance with the setting range
stated in this
IFU
.
(See Fig. 5)
11 Adjust water temperature so that the disin-
fectant mixtures reaches approximately
20° C (68°F).
12 Spray the lift hygiene chair with disinfect-
ant.
(See Fig. 6)
13 Use disposable cloths or paper towels to
scrub the chair surfaces. Make sure under-
side of lift hygiene chair seat is being disin-
fected as well.
14 Close the disinfection control and put back
the disinfection spray handle.
15 Allow a disinfection time according the dis-
infectant instructions.
16 Grab the shower handle and open the
shower control.
17 Rinse with cold water until there are no
visible remains of the disinfectant.
18 Use the flush function if the
Shower &
Disinfection panel
is equipped with a flush
valve or a flush button.
19 Wipe the lift hygiene chair and the
Prelude
dry with a clean cloth or towel to remove
excess water.
Содержание Prelude
Страница 1: ... with people in mind Prelude 04 BB 01_6US CA June 2012 INSTRUCTIONS FOR USE ...
Страница 9: ...9 Model without toilet function Model with toilet function 1 1 2 2 3 3 4 4 5 6 7 5 6 8 9 10 11 8 7 ...
Страница 11: ...11 1 2 3 4 ...
Страница 13: ...13 1 2 3 4 M X M X IX ...
Страница 15: ...15 1 2 3 4 5 6 X ...
Страница 17: ...17 1 2 3 4 5 ...
Страница 19: ...19 1 2 3 4 5 6 7 8 ...
Страница 23: ...23 1 2 3 ArjoHuntleigh Arjo C ean Arjo Clean 50 ...
Страница 25: ...25 1 2 3 4 5 6 Arjo Clean 50 ArjoHuntleigh Arjo Clean ...
Страница 27: ...27 1 2 3 4 5 6 Arjo Clean 50 A joHunt eigh Arjo Clean ...
Страница 31: ...31 1 2 Arjo Clean 50 M X ...
Страница 37: ...37 Intentionally left blank ...
Страница 38: ...38 Intentionally left blank ...