15
4.3.2 Service
table
Note!
The service table is only shown for information purposes.
WARNING!
Maintenance may only be performed by authorized
service technicians.
WARNING!
The machine is connected to the electricity supply
and some components are live.
The service table below shows recommended maintenance
intervals.
In addition to system testing, we recommend that appropriate
cleaning tests and temperature validation are performed according
to EN ISO 15883.
Measure
Every
year/ 7,500
cycles
Every other
year/15,000
cycles
Time*
(minutes)
1
General
1.1
Check the cabling and connection points.
•
10
1.2
Check the function of the cooling fan. Clean as required.
•
5
1.3
&KHFNWKDWWKHSDQHOODEHOLV¿UPO\DI¿[HG
is undamaged and legible.
•
2
1.4
Check that the door lock and breaker are functioning
properly.
•
5
1.5
Check that a goods placement sign and quick help
instructions have been put up.
•
1
1.6
Check error codes and the number of processes.
•
2
2
The chamber
2.1
&KHFNWKDWWKHURWDWLQJQR]]OHVDUHPRYLQJIUHHO\
Clean as required.
•
20
2.2
&KHFNWKDWWKH¿[HGQR]]OHVDUHQRWFORJJHGZLWK
grime and deposits. Clean as required.
•
20
2.3
&KHFNIRUOHDNDJHLQWKHQR]]OHDWWDFKPHQWVDQGWKH
hose connections to the chamber.
•
-
2.4
Check for proper operation of the temperature sensor.
•
10
2.5
Check that the door seal and chamber seal are tight.
Clean or replace as necessary.
•
10
2.6
Check the mounting for the goods holder.
•
2