17
00
1-
00
55
3-
EN
r
e
v
.
5
condition on a meter.
Battery Test
Battery without fuse - former labeling
With former battery tester.
1)
The battery to be tested must be charged for at least 12 hours before the test.
2)
Connect the tester to the battery. When the tester is fully connected, turn the test switch right or
left for 1-2 seconds and take a quick reading on the meter.
3)
The meter will show the battery’s condition.
.
Battery with fuse - new labeling
With battery tester (refer to SEE201204).
1)
The battery to be tested must be charged for at least 12 hours before the test.
2)
Connect the tester to the battery. When the tester is fully connected, press and hold the button
for 1-2 seconds and take a quick reading on the meter.
3)
The meter will show the battery’s condition.
NOTE: The battery power tester will become hot and there is a potential risk of fire.
• Make sure the battery power tester is kept away from flammable
material during test and for several minutes AFTER the test.
• Beware NOT TO TOUCH the metal casing for several minutes AFTER the test.
Tester (8662950) for battery with fuse
Tester for battery without fuse
Fig. 9