40
Every Month
•
Test the thermostatic mixer
Turn the temperature control back and forth over the entire
temperature range with water on. This is particularly
important in areas with hard water.
(See Fig. 1)
•
Check/clean filters for incoming water:
•
A shut-off valve with a filter is mounted on each of the two
water supply hoses found under the front cover. The filter in
both valves should be rinsed as follows:
1 Close the valves
A
). The handles (red for hot water and
blue for cold) should be in a cross position.
(See Fig. 2)
2 Remove the cap (
B
) with a wrench.
3 Pull out the filter (
C
) and clean it carefully in water.
4 Reassemble.
5 Open the valves.
•
Check expiring date for seat, replace before expired:
Look at the label located on the back of the seat to find the
expiration date.
(See Fig. 3)
WARNING
Pressurised system and risk of scalding. Make
sure that the valve (A) is closed before removing
the cap, this is especially important for the hot
water valve.
Fig. 1
Fig. 2
A
B
C
Fig. 3
Expiration date of the seat