37
Care and Preventive Maintenance
The
Malibu/Sovereign
bath is subject to wear and tear, and the following actions must be per-
formed when specified to make sure that the product remains within its original manufacturing
specification.
PREVENTIVE MAINTENANCE SCHEDULE: Malibu/Sovereign bath
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow
the recommended maintenance schedule. In some cases due to heavy use of the product and
exposure to aggressive environment more frequent inspections should be carried out. Local
regulations and standards may be more stringent than the recommended maintenance
schedule.
NOTE
Product cannot be maintained and serviced while in use with the patient.
CAREGIVER OBLIGATIONS Action/Check
Between
Patients
Every
Week
Every
Month
Every
Year
Every
2:nd
Year
Clean and Disinfect
X
Visually check all exposed parts
X
Visually check mechanical attachments
X
Visually check hoses, pipes and
connections
X
Check hand control and cable
X
Check castors on wheel chassis
X
Perform functionality test
X
Check neck pillow
X
Visually check the safety belt
X
Exercise thermostat mixer
X
Check/clean filters, incoming water
X
Check expiring date for seat, replace before
expired
X
Replace the safety belt
X
Yearly checks by qualified personnel only
X
Check electrical installation (to be
performed by authorized electrician)
Intervals according to local requirements
WARNING
To avoid injury to both the patient and the caregiver, never modify the equipment or use
incompatible parts.
Содержание Malibu
Страница 1: ... with people in mind 04 AZ 00_14GB 12 2018 Malibu Sovereign Instructions for Use ...
Страница 47: ...47 Intentionally left blank ...
Страница 51: ...51 Intentionally left blank ...
Страница 56: ...Intentionally left blank ...
Страница 57: ...Intentionally left blank ...
Страница 58: ...Intentionally left blank ...