Electrical Installation
•
First check that the electrical supply corresponds
with the information on the data label.
•
The unit must be permanently installed.
•
The unit must be continuously powered.
The unit shall be electrically connected with the
following equipment (not included in delivery):
•
Separate fuse.
•
Ground fault circuit interrupter (10mA)
•
Safety switch, according to regulations and local
codes.
•
Mains cable (3x16AWG / 3x1.5 mm
2
)
Connect the wires as described below, Europe:
•
Neutral to the blue wire (N)
•
Phase to the brown wire (L)
•
Protective earth to the yellow/green wire (PE)
Connect the wires as described below,
US and Canada:
•
Neutral to the white wire (N)
•
Phase to the black wire (L)
•
Protective earth to the green wire (PE)
The unit must be permanently connected to the mains
when used in USA and Canada. All electrical installa-
tions must comply with the US National Electrical Code
and the Canadian Electrical Code respectively.
Equapotential bonding connection
Connect the equapotential bonding cable for accessing
the equapotential bonding point to the unit, open the
service cover to the right of the tub.