If the result of any of these tests is unsatisfactory, contact an Arjo
Huntleigh approved service agent.
WARNING
The procedures below must be carried out by suitably trained
and qualified personnel. Failure to do so may result in injury
or an unsafe product.
Actions to be done by qualified personnel
Daily
Weekly
Yearly
Carry out a full test of all electrical bed
positioning functions (backrest, height, tilt,
etc.)
Ask a person of known weight to lie on the
bed and check that their weight is indicated
correctly on the weighing system display
Enable the patient egress detection system
and check that the patient egress alarm is
triggered when the person gets off the bed
Check that the bed operates correctly when
using battery power- see below
Check operation of the manual CPR release
levers on both sides of the bed
Check operation of castors, paying particular
attention to braking and steering functions
Check that the bed extension locks securely
in both the normal and extended positions
Check that the bedstripper
latches securely
in the closed position
Examine the electrical supply cable and plug;
replace the complete assembly if damaged,
do not use a rewireable plug or adapter
Examine all accessible flexible cables for
damage and deterioration
Check all nuts, bolts and other fasteners are
present and correctly tightened
Check all bed accessories, paying particular
attention to fasteners and moving parts