27
746-449-SA-5
Additional ACP features
Function
lockout
Under certain circumstances it may be necessary to
prevent operation of the bed controls, e.g. when traction
equipment is being used.
To selectively disable bed functions:
Press the
Function Lockout
key on the Attendant Control
Panel. The indicator above the
Lockout
key will light red.
Press the ACP Function key corresponding to the function
to be disabled. The indicator above the key will light
yellow.
To disable ALL bed functions simultaneously:
Press and hold the Function Lockout key for five seconds.
If no Function key is pressed within five seconds of pressing
the Lockout key, the indicator above the Lockout key will go
out and the Lockout function must be selected again.
Function Lockout settings are retained when the bed is
disconnected from the electricity supply.
To reinstate previously disabled bed functions:
Press the
Function Lockout
key on the Attendant Control
Panel. The indicator above the Lockout key will light.
Press the ACP Function key corresponding to the function
to be reinstated. The indicator above the key will go out.
CPR emergency
Press and hold the
CPR
key. This will flatten, and if
necessary lower, the mattress platform to enable cardio-
pulmonary resuscitation to be carried out.