6.  SORRENTO SINGLE BED SPECIFICATIONS

 

 

  Basic Dimensions 

 

Overall Length 

211.5cm 

Overall Length (base extension extended) 

231.0cm 

Overall Width 

92.0cm 

Overall Width (rails fitted) 

102.5cm 

Mattress Surface Length (between panels) 

207.0cm 

Bed Platform Height Range 

17.0cm to 69.0cm (+/- 5mm) 

Back Rest Angle Adjustment 

0° to 70° 

Knee Bend Angle Adjustment 

0° to 35° 

Trendelenberg Tilt Adjustment (head and foot down)  

0° to 14° 

 

  

Mattress Size 

 

Recommended Mattress Size 

Width   = 86.0cm  

Length = 198.0cm     

Maximum Mattress Depth  17.0cm 

Minimum Mattress Depth  10.0cm 

 

  

Bed Weight & Safe Working Load

 

 

Without Side Rails 

119 Kg (approximately) 

With Side Rails 

138 Kg (approximately) 

Safe Working Load Safety Factor = 2 x S.W.L  

250 Kg 

Maximum Patient Weight 

215 Kg 

 

  Electronic Parameters

 

 

Power Input 

100-240V AC 50/60Hz

Max current 3.9 A 

Rate Load 

interruption 10%, Max 2 minutes / 18 minutes 

Electric Shock Protection 

 

Waterproof Grade 

IPX4 

 

  

Standards 

 

Applicable standards to the Sorrento bed 

EN 980:2009, EN 1041:2008, EN ISO 14971:2019, IEC 60601-1:2005+A1:2012, IEC 60601-1-
2:2014, IEC 60601-2-52:2009+A1:2015, ISO 10993-1:2018, ISO 13485:2016 

The Sorrento bed is built to a specialty and deviates in clauses (201.9.2.2 Trapping Zone, 
201.9.2.2.1 General) in standard IEC 60601-2-52:2009+A1:2015 to provide the low height 
mattress support platform that helps improve patient safety from falls. The deviations have 
been benefit/risk assessed in the risk management processes, taking guidance and review 
from Subclause 201.9.1 Mechanical Hazards of ME Equipment. The bed has been recognised 
as suitable for market use.

 

 

Содержание Sorrento Floor Bed

Страница 1: ...Instructions for Use Single King Single Bed Options Sorrento Floor Bed SOR REV5 090921...

Страница 2: ...opulmonary Resuscitation CPR 8 4 9 Lock Out Function 8 4 10 Battery Charge Indicator 8 4 11 Mains Power Indicator 9 4 12 Under Bed Light 9 5 GENERAL BED OPERATION 10 14 5 1 Castor System 10 5 2 Bed Ex...

Страница 3: ...iate for long term adult patient care under supervised medical monitoring and provides support with various positioning options Key Features Total safe working load SWL of the bed 250Kg Four 75mm brak...

Страница 4: ...equipment can be hazardous if used incorrectly Do not use electrically operated beds in the presence of flammable gases Use the bed to support one patient only The Sorrento bed is intended for one pe...

Страница 5: ...eft in the flat position whilst unattended The clinically qualified person responsible should consider the age size and condition of the patient before using the bed and its accessories Always ensure...

Страница 6: ...Use Revision SOR REV5 090921 3 ELECTRONIC FUNCTIONS 3 1 Handset Foot Board Control Panel Operation The Sorrento s bed adjustments are powered by four electric actuators which are controlled by the but...

Страница 7: ...ctuator force in the event of being obstructed in lowering 4 2 Mattress Platform Height Adjustment WARNING It is recommended that the bed be positioned at its lowest height when unattended by caregive...

Страница 8: ...t or head down tilt position Tilt positioning achievable is between 0 and 14 for both head and foot down tilting In use the mattress platform may initially rise before tilting The bed will automatical...

Страница 9: ...on the control panel by LED lights in between the function buttons To lock and unlock each function press the lock button and at the same time press the function button that is required to be locked...

Страница 10: ...l devices fitted to the bed The under bed lighting feature provides a soft LED illumination of the floor beneath the bed carriage This safety feature provides floor illumination for patients at night...

Страница 11: ...ase the braking castors use the pedals at each side and end of the bed Simply push down on the red lever using the toe of the shoe to brake and the green lever to release the brake CAUTION Disconnect...

Страница 12: ...screw the handwheels both sides of the bed anti clockwise until approximately 20mm of thread is showing Pull the foot panel towards you using two hands evenly spaced on the panel Pull the panel to its...

Страница 13: ...et these requirements when fitted correctly 5 4 Fitting the Side Rails NOTE The rails are a left and right hand set and when fitted correctly fold down towards the foot end of the bed 1 Before fitting...

Страница 14: ...rs visitors and or other equipment Side Rail Operation To raise the side Lift the top rail until the magenta locating pin knob automatically clicks into its locked position at maximum height To lower...

Страница 15: ...ellow shaft visible If there is yellow shaft visible the safety side is NOT securely locked in the upright position and needs to be adjusted The yellow coating should NOT be visible from above when th...

Страница 16: ...Safe Working Load Safety Factor 2 x S W L 250 Kg Maximum Patient Weight 215 Kg Electronic Parameters Power Input 100 240V AC 50 60Hz Max current 3 9 A Rate Load interruption 10 Max 2 minutes 18 minute...

Страница 17: ...ely Safe Working Load Safety Factor 2 x S W L 250 Kg Maximum Patient Weight 215 Kg Electronic Parameters Power Input 100 240V AC 50 60Hz Max current 3 9 A Rate Load interruption 10 Max 2 minutes 18 mi...

Страница 18: ...hot water and mild detergent or the facilities recommended cleaning solution It is recommended that the cloth be a white material to prevent bleeding of die colours to the bed Take extra care in areas...

Страница 19: ...ASTOR BRAKING LOCKING SYSTEM 1 Apply the brakes and push the bed forwards If any of the four braking castors rotate the brake is not fully effective SIDE RAILS if fitted 1 The locking pin mechanism an...

Страница 20: ...dset is activated Control box requires calibrating possibly due to bed being operated when batteries were fully discharged Ensure mains power lead is connected and carry out calibration procedure Brak...

Страница 21: ...he product must be recycled separately Batteries are to be disposed in accordance with national or local regulations Components that are primarily made up of different kinds of metal containing more t...

Страница 22: ...Page 21 Commercial in confidence Instructions For Use Revision SOR REV5 090921 14 SORRENTO BED LABELS WARNINGS...

Страница 23: ...abel Areas labelled are of a low risk nature Recommended Mattress Size Label Manufacturer European Representative Date of Manufacture Product Code Reference Serial Number Bed Safe Working Load Maximum...

Страница 24: ...ions which provides the full details of warranty terms and statutory rights of the customer can be obtained upon request 17 REGULATORY INFORMATION Caretek China Medical PLC Xiguakeng Guanqiao Village...

Отзывы: