10.  TROUBLE SHOOTING

 

 

Symptom 

Possible Cause 

Action 

 
All actuators fail to work 

Power disconnected. or 
batteries discharged 

Ensure mains supply lead is 
connected 

 

Control box and handset 
plugs disconnected 

Check the plugs are fully 
inserted in the control box 
sockets. 

 

Blown fuse 

Contact your local Arjo 
Service Department. 

 

Damaged handset cable 

Replace handset 

One actuator fails to work 

Actuator plug disconnected 

Check that actuator plug is 
fully inserted in control box. 

Main lift actuators fail to work, 
and an audible beeping sound 
can be heard when handset is 
activated 

Control box requires 
calibrating possibly due to bed 
being operated when batteries 
were fully discharged 

Ensure mains power lead is 
connected and carry out 
calibration procedure. 

Brake lock not effective 

Wear or damage 

Adjust or replace castor 

 

11.  TRANSPORTATION AND STORAGE

 

 

Handle With Care. 

 
The Sorrento bed should be stored in a clean, dry, well ventilated area. 

Do not drop and avoid shock, or violent impact when transporting.  

The following limits apply during transport and during a storage period of up to 15 weeks 
duration: 

Ambient temperature 

-25°C 

To 

+70°C 

Relative humidity 

10% 

To 

75% 

Air pressure 

50 kPa   

To 

106 kPa 

 

The following limits apply to normal operating conditions and for periods of storage longer 
than 15 weeks:

 

Ambient temperature 

+10°C 

To 

+40°C 

Relative humidity 

30% 

To 

75% 

Air pressure 

70 kPa   

To 

106 kPa 

 

 

Содержание Sorrento Floor Bed

Страница 1: ...Instructions for Use Single King Single Bed Options Sorrento Floor Bed SOR REV5 090921...

Страница 2: ...opulmonary Resuscitation CPR 8 4 9 Lock Out Function 8 4 10 Battery Charge Indicator 8 4 11 Mains Power Indicator 9 4 12 Under Bed Light 9 5 GENERAL BED OPERATION 10 14 5 1 Castor System 10 5 2 Bed Ex...

Страница 3: ...iate for long term adult patient care under supervised medical monitoring and provides support with various positioning options Key Features Total safe working load SWL of the bed 250Kg Four 75mm brak...

Страница 4: ...equipment can be hazardous if used incorrectly Do not use electrically operated beds in the presence of flammable gases Use the bed to support one patient only The Sorrento bed is intended for one pe...

Страница 5: ...eft in the flat position whilst unattended The clinically qualified person responsible should consider the age size and condition of the patient before using the bed and its accessories Always ensure...

Страница 6: ...Use Revision SOR REV5 090921 3 ELECTRONIC FUNCTIONS 3 1 Handset Foot Board Control Panel Operation The Sorrento s bed adjustments are powered by four electric actuators which are controlled by the but...

Страница 7: ...ctuator force in the event of being obstructed in lowering 4 2 Mattress Platform Height Adjustment WARNING It is recommended that the bed be positioned at its lowest height when unattended by caregive...

Страница 8: ...t or head down tilt position Tilt positioning achievable is between 0 and 14 for both head and foot down tilting In use the mattress platform may initially rise before tilting The bed will automatical...

Страница 9: ...on the control panel by LED lights in between the function buttons To lock and unlock each function press the lock button and at the same time press the function button that is required to be locked...

Страница 10: ...l devices fitted to the bed The under bed lighting feature provides a soft LED illumination of the floor beneath the bed carriage This safety feature provides floor illumination for patients at night...

Страница 11: ...ase the braking castors use the pedals at each side and end of the bed Simply push down on the red lever using the toe of the shoe to brake and the green lever to release the brake CAUTION Disconnect...

Страница 12: ...screw the handwheels both sides of the bed anti clockwise until approximately 20mm of thread is showing Pull the foot panel towards you using two hands evenly spaced on the panel Pull the panel to its...

Страница 13: ...et these requirements when fitted correctly 5 4 Fitting the Side Rails NOTE The rails are a left and right hand set and when fitted correctly fold down towards the foot end of the bed 1 Before fitting...

Страница 14: ...rs visitors and or other equipment Side Rail Operation To raise the side Lift the top rail until the magenta locating pin knob automatically clicks into its locked position at maximum height To lower...

Страница 15: ...ellow shaft visible If there is yellow shaft visible the safety side is NOT securely locked in the upright position and needs to be adjusted The yellow coating should NOT be visible from above when th...

Страница 16: ...Safe Working Load Safety Factor 2 x S W L 250 Kg Maximum Patient Weight 215 Kg Electronic Parameters Power Input 100 240V AC 50 60Hz Max current 3 9 A Rate Load interruption 10 Max 2 minutes 18 minute...

Страница 17: ...ely Safe Working Load Safety Factor 2 x S W L 250 Kg Maximum Patient Weight 215 Kg Electronic Parameters Power Input 100 240V AC 50 60Hz Max current 3 9 A Rate Load interruption 10 Max 2 minutes 18 mi...

Страница 18: ...hot water and mild detergent or the facilities recommended cleaning solution It is recommended that the cloth be a white material to prevent bleeding of die colours to the bed Take extra care in areas...

Страница 19: ...ASTOR BRAKING LOCKING SYSTEM 1 Apply the brakes and push the bed forwards If any of the four braking castors rotate the brake is not fully effective SIDE RAILS if fitted 1 The locking pin mechanism an...

Страница 20: ...dset is activated Control box requires calibrating possibly due to bed being operated when batteries were fully discharged Ensure mains power lead is connected and carry out calibration procedure Brak...

Страница 21: ...he product must be recycled separately Batteries are to be disposed in accordance with national or local regulations Components that are primarily made up of different kinds of metal containing more t...

Страница 22: ...Page 21 Commercial in confidence Instructions For Use Revision SOR REV5 090921 14 SORRENTO BED LABELS WARNINGS...

Страница 23: ...abel Areas labelled are of a low risk nature Recommended Mattress Size Label Manufacturer European Representative Date of Manufacture Product Code Reference Serial Number Bed Safe Working Load Maximum...

Страница 24: ...ions which provides the full details of warranty terms and statutory rights of the customer can be obtained upon request 17 REGULATORY INFORMATION Caretek China Medical PLC Xiguakeng Guanqiao Village...

Отзывы: