32
Electromagnetic Compatibility
Conducted
disturbances
inducted by RF
fields
EN 61000-4-6
3V in 0,15 MHz to 80 MHz
6V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz
80% AM at 1 kHz
3V in 0,15 MHz to 80 MHz
6V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz
80% AM at 1 kHz
Portable and mobile RF
communications
equipment should be used
no closer to any part of
the product, including
cables, than 1.0m, if the
transmitter’s output
power rating exceeds
1W
a
Field strengths from
fixed RF transmitters, as
determined by an
electromagnetic site
survey, should be less
than the compliance level
in each frequency range
b
Interference may occur in
the vicinity of equipment
marked with this symbol:
Radiated RF
electromagnetic
field
EN 61000-4-3
Home Healthcare environment
10 V/m
80 MHz to 2,7 GHz
80% AM at 1 kHz
Home Healthcare environment
10 V/m
80 MHz to 2,7 GHz
80% AM at 1 kHz
Proximity fields
from RF wireless
communications
equipment
EN 61000-4-3
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz - 9V/m
810, 870, 930 MHz - 28 V/m
1720, 1845, 1970, 2450 MHz –
28 V/m
5240,5500, 5785 MHz - 9V/m
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz - 9V/m
810, 870, 930 MHz - 28 V/m
1720, 1845, 1970, 2450 MHz –
28 V/m
5240,5500, 5785 MHz - 9V/m
Electrical fast
transient/burst
EN 61000-4-4
±1kV SIP/SOP ports
100 kHz repetition frequency
±1kV SIP/SOP ports
100 kHz repetition frequency
Power frequency
Magnetic field
EN 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the product is used exceeds
the applicable RF compliance level above, the product should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Caution:
Operating the product near a strong electro-magnetic field may affect the displayed weight
measurement.
These changes does not affect user safety.