background image

19

A battery that is charged for the first time, or after a long 

storage period, must be charged until the charger 

indicates full charge.

Battery discharge indicator

The 

Sara Plus

 incorporates a Battery discharge indicator, 

situated on the right hand side of the cover. 

(See Fig 2 on page 7 in section “Parts Designation”)

The display shows eight levels of battery state ranging 

from fully charged on the right to very low on the left 

(filled segment to empty segment).

It is recommended that the battery is removed from the 

lift and charged when the display reaches the 3 filled 

segments and the buzzer beeps once every 10 seconds. 

Lifting is possible until the display shows one filled 

segment and the buzzer beeps continuously. At this 

point, the battery must be charged as soon as possible.

Warning:

 The charging of the battery must 

only be performed away from the patient 

environment.
The charger is for indoor use only.
Only use the charger unit in a dry environment, 

do not use it in the bathroom.
Do not expose the charger unit or battery pack 

to rain or spray and do not immerse in water.
Do not expose the charger unit to dust.
To avoid overheating, the charger must not be 

covered whilst in use.
No smoking or naked flames in battery vicinity.
The battery charger is for use only with Arjo 

supplied batteries that are to be used with the 

Sara Plus

.

The battery charger is for use with sealed lead 

acid batteries only.
Under no circumstances should the charger be 

used to attempt to charge non-rechargeable 

batteries.
Do not attempt to open or tamper with the 

charger unit in any way, for any repair the 

charger must be sent to the manufacturer.
The mains electricity socket must be easily 

accessible. Should a faulty condition occur 

switch off and remove the connection plug 

from the socket.
Only use Arjo components that have been 

specifically designed for the purpose when 

charging batteries
Only use the Arjo battery that is supplied to be 

used with the 

Sara Plus

Only use the Arjo charger unit supplied with 

the 

Sara Plus

.

Do 

NOT

 place batteries near, or dispose of, in 

a fire.
Do 

NOT

 short circuit a battery.

Do 

NOT

 store batteries at temperatures in 

excess of 60°C (140°F).

Warning:

Do 

NOT

 crush, puncture, open, 

dismantle or otherwise mechanically interfere 

with batteries.
Should the battery casing become cracked, and 

electrolyte come into contact with skin or 

clothing, wash immediately with water.
If the electrolyte contacts the eyes, wash 

immediately with copious amounts of water, 

and seek medical attention.
When disposing of batteries, contact the 

appropriate local authority for advice.
The abbreviation “Pb” shown adjacent to the 

recycling and trash bin symbols on the battery 

back label is the element symbol for lead, and 

indicates that the battery contains lead and 

therefore should not be disposed of in the 

normal manner but must be recycled.

Make sure the battery is removed from the lift 

if it is anticipated it will not be used for a 

prolonged period of time.

Note:

The battery discharge indicator has an energy 

saving function, automatically switching off 

the display if a function button has not been 

operated for at least 60 seconds. The moment 

a button is pressed to operate any function, the 

display will re-start.

Note:

Battery Instructions

Содержание SARA PLUS

Страница 1: ...Sara Plus KKX52180M EN_21 0 2020 INSTRUCTIONS FOR USE...

Страница 2: ...s improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either whole or in part without the consent of Arjo WARNING To avoid injury...

Страница 3: ...s legs 10 Scale if available 10 Allowed slings with Sara Plus 10 Commode seat Accessory 11 How to use the Sara Plus 12 General 12 Prepare Sara Plus and patient 12 Use the Standing sling for transfer 1...

Страница 4: ...t occur as a result of any unauthorized modification to its products Support If you require further support please contact Arjo for comprehensive support and maintenance to maximize the long term safe...

Страница 5: ...eds mobility maintaining standing exercises When used in combination with an EPS Extra Postural Support BOS sling Buttocks support standing sling and only for raising and transferring purpose the Sara...

Страница 6: ...is 190kg 420lbs Do not exceed these weight limits Warning It is advisable to familiarise yourself and understand the operation of the various controls and features of the Sara Plus as described in Pro...

Страница 7: ...use 4 steps 1 Visually inspect Sara Plus If any part is damaged do NOT use the product 2 Check the battery level 3 Make sure the green reset button situated on the rear of the mast is pressed i See Fi...

Страница 8: ...3 Dual control panel 14 Hand control 15 System failure lower override knob 16 Label System failure lower override identification 17 Handset cable connection 18 Battery discharge indicator 19 Power on...

Страница 9: ...nadvertently overloaded trying to lift a patient heavier than permitted an automatic cut out operates to prevent the lift raising a load in excess of the safe working load this stops the lift motion a...

Страница 10: ...ce to follow the intended straight line walked by the patient The function is activated by flipping over the steering guide on the rear castor to hold it in position See Fig 6 Arc Rest with handgrips...

Страница 11: ...opened to any variable width See Fig 10 When pressure is released from the button movement stops and the chassis legs remains securely in position Scale if available If your Sara Plus has been suppli...

Страница 12: ...patient has been lifted to a standing or near standing position in the manner previously described See Fig 11 Removal of any clothing can be attended to and the patient is then lowered down onto the c...

Страница 13: ...lowered sufficiently to avoid approaching the patient at eye level and making allowances for the patients arms and any obstructions e g chair arms etc 3 Approach the patient from the front with the li...

Страница 14: ...e attached correctly make adjustments on both cords equally so that any slack is taken up in each cord and the back section of the sling supports the patient comfortably and securely Fig 16 Warning As...

Страница 15: ...nger distance transportation See Fig 21 The patient should be supported by the sling but not pulled forward too much See Fig 18 Note Fig 18 Detail view showing cords cleats Fig 19 If the patient can o...

Страница 16: ...trap repeat for the other side See Fig 22 and Fig 17 for attachment of the cords 2 When both cords are attached correctly make adjustments on both cords equally so that any slack in the cord is taken...

Страница 17: ...es and close the chassis legs then transport the patient to desired position See Fig 26 13 Lower the patient carefully using the hand control or dual control panel 14 When the patient is seated in the...

Страница 18: ...connect the attachment clips to the lugs on the outer sides of the Arc Rest See Fig 28 5 When the patient is ready operate the lift button on the handset or dual control to raise the patient at the s...

Страница 19: ...available refer to the Scale IFU Fig 31 Warning When the Proactive Pad assembly is removed make sure that the attachment bracket is also removed before starting therapy Failure to do so could lead to...

Страница 20: ...empt to open or tamper with the charger unit in any way for any repair the charger must be sent to the manufacturer The mains electricity socket must be easily accessible Should a faulty condition occ...

Страница 21: ...arged This will change to a yellow light as the battery approaches full charge capacity finally changing to a green light when the battery is fully charged 5 Charge the battery until the charger indic...

Страница 22: ...pment needs cleaning or are suspected of being contaminated follow the cleaning and or disinfection procedures recommended below before re using the equipment This is especially important when using t...

Страница 23: ...Commode seat 2 steps 1 Pull the rear edge up sharply to disengage the locating lugs 2 Slide the seat forwards a short distance until clear of the frame tubes and lift away See Fig 34 Attach the Commod...

Страница 24: ...ors Replace as required X S Check that the covers fit correctly and are not damaged Replace as required X S Check for evidence of corrosion Replace as necessary X S Warning The points on this checklis...

Страница 25: ...n in the upper lift arm is tightened Make sure the screws retaining the cord cleats in the Arc Rest are tight Make sure all instruction labels are firmly attached and are readable Examine all exposed...

Страница 26: ...y an obstruction Remove the obstruction and check the Sara Plus thoroughly for damage before continuing the lifting cycle If in doubt use the System failure lower override to return the resident to a...

Страница 27: ...SPL3021 Battery capacity 6 Ah Battery charger part number Note indicates relevant country code SPL3024 Fuse 15 A Thermal overcurrent circuit breaker Fuse PCBA 20 A Fuse battery 30 A Protection class...

Страница 28: ...and Storage Atmospheric pressure 800 hPa to 1060 hPa Operating 500 hPa to 1100 hPa Transport 500 hPa to 1100 hPa Storage End of Life Disposal All batteries in the product must be recycled separately B...

Страница 29: ...28 Dimensions Sara Plus dimensions Dimensions in millimetres equivalent in inches...

Страница 30: ...entification 5 Address and SWL 190 kg 420 lbs 6 Product name 7 Safe working load 190 kg 420 lbs 8 CE mark 9 System failure lower override identification 10 Sling size guide 11 Read operating instructi...

Страница 31: ...duct is protected against insertion of fingers and splashing water Type BF Applied part protection against electrical shockinaccordancewith EN IEC 60601 1 CE marking indicating conformity with Europea...

Страница 32: ...ence in nearby electronic equipment RF emissions CISPR 11 Class B This equipment is suitable for use in all establishments including domestic establishments and those directly connected to the public...

Страница 33: ...z 28 V m 710 745 780 MHz 9V m 810 870 930 MHz 28 V m 1720 1845 1970 2450 MHz 28 V m 5240 5500 5785 MHz 9V m Electrical fast transient burst EN 61000 4 4 1kV SIP SOP ports 100 kHz repetition frequency...

Страница 34: ...Intentionally left blank...

Страница 35: ...22 E mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Of ce 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 48...

Страница 36: ...e ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to cont...

Отзывы: