5. CLEANING
Cleaning
It is strongly recommended that protective clothing be used when carrying
out any cleaning processes on the Prioma Overbed Table.
•
Wipe the Overbed Table with a soft cloth moistened with hot water and mild detergent (or
hospital’s recommended cleaning solution). It is recommended that the cloth be a white
material to prevent bleeding of die colours onto the product surface. Take extra care in
areas that can harbour dirt or dust.
•
Start by cleaning the upper sections of the Overbed Table and work along all horizontal
surfaces. Work methodically towards the lower areas and clean the castors last. Take
extra care to clean areas that may trap dust or dirt.
•
Rinse with clean water and a dry clean cloth.
•
To remove potentially infectious materials such as body fluids, or when the product has
been used by someone with a known infection. Clean with NaDCC (e.g. Presept,
Actichlor) at 10,000 ppm of available chlorine. Rinse and dry.
•
Allow the cleaned parts to dry before replacing the mattress.
Caution
Do not use abrasive Hypo-carbonate or Phenolic compounds or pads.
Disinfecting
After cleaning the Overbed Table as described above, wipe all surfaces with sodium
dichloroisocyanurate (NaDCC) at a concentration of 1,000 parts per million (0.1%) of
available chlorine.
In the case of pooling body fluids, e.g. blood, the concentration of NaDCC should be
increased to 10,000 parts per million (1%) of available chlorine.
Caution
Avoid excessive wetting of the product, particularly the tabletop.
Do not use abrasive compounds or pads.
Do not use phenol-based disinfectants.
Do not use excessively hot water. The tabletop has a maximum heat resistance
temperature of 60°C or 140°F.