background image

Service Procedures

MRM Carendo

4

Service procedures not included in PMS

Right side

Left side

SP 13

 Replace 

worm gear/motor
p.18

SP 14

 Replace 

actuator p.20

SP 12

 

Replace 

locking device 
p.17

SP 15 

Replace 

emergency 
buttons p.23

SP 11

 

Replace 

PCBA and 
fuses p.16

Содержание CARENDO

Страница 1: ...09 CC 01_8EN 01 2019 MAINTENANCE AND REPAIR MANUAL Carendo For internal use only...

Страница 2: ...4 Lubrication Sealants Lockings and Torques 24 Tools 30 Trouble shooting 31 NOTE Before using this manual you must have taken part of and understood the Safety Precautions 09 NE 01 Blue text indicate...

Страница 3: ...chanical attachments x 4 8 Lubricate bearings x 5 9 Perform a safe working load test x 6 10 Check the safety stop function x 7 11 Check function of emergency lowering x 8 12 Perform full feature funct...

Страница 4: ...4 Service procedures not included in PMS Right side Left side SP 13 Replace worm gear motor p 18 SP 14 Replace actuator p 20 SP 12 Replace locking device p 17 SP 15 Replace emergency buttons p 23 SP...

Страница 5: ...n a Field Safety Corrective Action Technical Advisory Notices are divided in three classes A Field Safety Corrective Action FSCA Information about what technical interventions to be made in order to r...

Страница 6: ...fore test 2 Connect the tester to the battery with suitable connector a 3 Press and hold button b for 1 2 seconds 4 The meter c will show the battery condition 5 Green indication PASS Yellow indicatio...

Страница 7: ...ion b Care raiser lubrication of locking pins see page 28 c Leg support and raiser d Seat e Back rest with storage net f Handset and cable g Safety strap h Buttons i Bellows j Castors See page SP 10 I...

Страница 8: ...on page 14 f Bellows strap or clamp g Buttons and handset cable h Locking device See also Lubrication Sealants Lockings and Torques on page 24 and Tools on page 30 NOTE Screws and bolts attached with...

Страница 9: ...in the slit and gently pulling up and outwards 5 Loosen the two screws c on the lower bearing 6 Be careful about the cable protection d on the right lower bearing 7 You can use the loose screws to pu...

Страница 10: ...edures MRM Carendo 10 SP 6 Perform a Safe Working Load Test Applicable load test Model Carendo SWL 136 kg 300 lb Test load 136 kg 300 lb General load test procedure is not ready Please refer to local...

Страница 11: ...t using both the handset and the emergency lowering at the same time holding an armrest to stop the motion 2 The lift shall stop 3 If not check cabling a from actuator to PCBA orange green lowering Ye...

Страница 12: ...Service Procedures MRM Carendo 12 SP 8 Check Function of Emergency Lowering 1 Press the button 2 Carendo shall lower 3 If not check button cabling to PCBA connection J7 fuse F3 before replacing PCBA...

Страница 13: ...Reclining of seat care raiser Handcontrol button Seat is reclining care raiser locked Raising of seat Handcontrol button Seat is raising Raising of seat care raiser Handcontrol button Seat is raising...

Страница 14: ...n of brakes if applicable by applying the brakes and push the product forward The brakes must lock the wheels and the brake lever must function on off 7 Apply load to the product approx 80 kg and do a...

Страница 15: ...o down 2 Loosen and remove the castor from the chassis 3 Clean the threads in the hole Refer to Loctite instructions 4 Apply Loctite see page 24 for quality 5 Assemble the new castor with new bolt and...

Страница 16: ...l not replace the PCBA New version PCBA from P0519014 Push UP on the hand control If not in lowest position also push the emergency lowering button If LED1 c on PCBA turn ON it indicates that fuse F1...

Страница 17: ...llen screws d 8 Lift out the complete mounting plate with the locking devices 9 Unscrew the two screws with nuts e on the locking device 10 Replace the locking device 11 When reassembling use new lock...

Страница 18: ...krest according to page 17 item 2 3 6 Support the backrest from below to prevent it from falling down 7 Use the loose screws on the bearing to pull out the bearings on both sides 8 Remove the care rai...

Страница 19: ...for the worm gear unit To replace only the motor 19 Loosen the two bolts and pull off the motor from the splines Reassembling 1 When reassembling notice the position of the cable 2 Use Loctite accord...

Страница 20: ...gs push couplings through hole of bearing 6 Turn the Carendo carefully on the side 7 Cut the top strap loosen the clamp b holding the bellows 8 Loosen the three Allen bolts c and pull out the cable d...

Страница 21: ...the actuator pins Old thick bracket illustrated 15 Stretch the cable a from outside 16 Attach the three Allen bolts c use Loctite according to page 24 17 Attach the bellow with a new clamp 18 If new...

Страница 22: ...Use Connectors NEW Cable 8451430 coloured cables Old Actuator Use Connection Block Old Cable 8451430 Old Actuator Use Connection Block Old Cable 8451430 NEW Actuator Use Connection Block NEW Actuator...

Страница 23: ...hen the cable with connector is loose unscrew the locking ring a for the button turning it counter clockwise Reassembling 4 When reassembling note that the peg in the button and in the connector shall...

Страница 24: ...43 on fitting bolts for actuator Loctite 243 on screws for bearing and lift arm Loctite 243 on screws for bearing and lift arm Loctite 243 on screws for bearing Loctite 243 on screws for bracket plate...

Страница 25: ...ket plate Torque 8 Nm OKS 403 on bearing chassis frame Loctite 243 on screws locking pins Loctite 2701 on shaft threads Loctite 243 on fitting bolts for actuator Loctite 243 on screws for bearing and...

Страница 26: ...ws for segments on Care raiser and seat Loctite 243 on screws for upper tie rod Loctite 243 on screws for bed pan holder Loctite 243 on screws for leg support Old care raiser up to SEE06020104 New car...

Страница 27: ...MRM Carendo Lubrication Sealants Lockings and Torques 27 New care raiser replaces old New care raiser uses new knob OKS 403 on bearing chair...

Страница 28: ...ld Soap water on o ring Wekem WS 267 on locking pins Loctite 243 on screws for locking pins Service cover up to SEE0347097 Old plate up to SEE0347097 Loctite 243 on plug Loctite 243 on plug Loctite 24...

Страница 29: ...43 on screws for locking pins Wekem WS 267 on locking pins New plate from SEE0347098 Service cover from SEE0347098 Loctite 243 on plug Loctite 243 on plug Soap water on o ring OKS403 on 4 screws for l...

Страница 30: ...Tools MRM Carendo 30 Tools For specification of tools please refer to the Parts List Service Articles and Special Tools...

Страница 31: ...ying PCBA J11 1 2 Electronic fault Check PCBA 11 Seat will not rise Fuse bad Check fuse F1 11 Contacts in wrong position Check contacts on PCBA J7 J9 J10 J11 Bad connection Check hand set and cables P...

Страница 32: ...ault Check replace locking device 12 Microswitch fault Check microswitch sitting PCBA J10 1 2 Electronic fault Check PCBA 11 Seat rises but lock for care raiser is not deac tivated Microswitch fault C...

Страница 33: ...PCBA 11 Chair will lift lower but makes noises Bad lubrication attachment wear Check bearings chassis lubricate replace 4 Bad attachment wear Check rods attachment 3 Bad lubrication attachment wear C...

Страница 34: ...e ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to cont...

Отзывы: