MRM Carendo
33
Chair
Chair will
not lift
Fuse bad
Check/replace fuse F1
11
Bad connection
Check hand set and cables
PCBA J:1
Connection fault
Check connection under
right cover
13
Actuator fault
Check/replace actuator
13
Electronic fault
Check/replace PCBA
11
Chair will
not lower
Fuse bad
Check/replace fuse F1
11
Bad connection
Check hand set and cables
-
Connection fault
Check connection under
right cover
13
Microswitch fault
Check microswitch down
PCBA
J2:2,3
Actuator fault
Check/replace actuator
13
Electronic fault
Check/replace PCBA
11
Chair will
lift/lower
but makes
noises
Bad lubrication/
attachment, wear
Check bearings chassis:
lubricate/replace
4
Bad attachment,
wear
Check rods; attachment,
3
Bad lubrication/
attachment, wear
Check bearings actuator:
lubricate/ replace
13
Bad attachment
Check yoke attachment
3
E
mer-
gency
down
E
mergency
down does
not work
Fuse bad
Check/replace fuse F3
11
Malfunction, bad
connection
Check/replace button and
cables to PCBA
14
Microswitch fault
Check microswitch,
emergency down
PCBA:
J2:6,7
Electronic fault
Check/replace PCBA
11
E
mer-
gency
stop
E
mergency
stop does
not work
Malfunction, bad
connection
Check/replace button and
cables to PCBA
11,14
Electronic fault
Check/replace PCBA
11
Part Problem
Cause
Action