MRM Carendo
Service Procedures
21
Reassembling
13.Attach the new actuator (use Loctite
according to page 24).
14.Push the cable (
a
) through the hole (use
some soap if it is difficult) and position the
actuator (do not forget the two plastic
bushings (
b
) on top of the actuator pins.
(Old thick bracket illustrated)
15.Stretch the cable (
a
) from outside.
16.Attach the three Allen bolts (
c
) (use Loctite
according to page 24).
17.Attach the bellow with a new clamp.
18.If new cabling with quick connectors: cut
off quick connectors (
d
)
19.Connect the cables to the connection block
(
e
) See page 23 for difference in connections
between old and new cabling.
20.Reassemble cover and battery
21.Check function
e
Old
18
b
a
New
d
14
a
c
b