![background image](http://html1.mh-extra.com/html/arjo/auralis/auralis_instructions-for-use-manual_2972279062.webp)
ELECTROMAGNETIC COMPATIBILITY (EMC)
62
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC IMMUNITY
The Auralis
SXPSLVLQWHQGHGIRUXVHLQWKHHOHFWURPDJQHWLFHQYLURQPHQWVSHFL¿HGEHORZ7KHFXVWRPHURU
the user of the Auralis
pump should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagentic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms 6Vrms ISM,
150KHz- 80MHz
80 % AM 1KHz
10 V/m
80 MHz to 2.7 GHz
10Vrms
150KHz- 2300MHz
10 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the pump, including cables,that
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
G ¥3
G ¥30+]WR0+]
G ¥30+]WR*+]
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
)LHOGVWUHQJWKVIURP¿[HG5)WUDQVPLWWHUV
as determined by an electromagnetic
site survey
a
, should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
DQGUHÀHFWLRQIURPVWUXFWXUHVREMHFWVDQGSHRSOH
a
)LHOGVWUHQJWKVIURP¿[HGWUDQVPLWWHUVVXFKDVEDVHVWDWLRQVIRUUDGLRFHOOXODUFRUGOHVVWHOHSKRQHVDQG
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
WKHRUHWLFDOO\ZLWKDFFXUDF\7RDVVHVVWKHHOHFWURPDJQHWLFHQYLURQPHQWGXHWR¿[HG5)WUDQVPLWWHUVDQ
HOHFWURPDJQHWLFVLWHVXUYH\VKRXOGEHFRQVLGHUHG,IWKHPHDVXUHG¿HOGVWUHQJWKLQWKHORFDWLRQLQZKLFK
the Auralis
pump is used exceeds the applicable RF compliance level above, the Auralis
pump should
be observed to verify normal operation. If abnormal operation is observed, additional measures may be
necessary, such as reorientating or relocating the Auralis
pump.
b
2YHUWKHIUHTXHQF\]WR0+]¿HOGVWUHQJWKVVKRXOGEHOHVVWKDQ9P