Figure 11
Figure 12
Figure 13
Figure 14
PRODUCT DESCRIPTION - AURALIS PUMP
19
detected. The indicator is shown for information only.
The pump continues to operate normally by using the
battery. There is no audible alarm.
Service indicator
The Service indicator (see Figure 11) indicates two
states:
• When the Auralis pump is ready for service,
the Service indicator shows yellow and the
front panel alarm indicators show green. This
is an information signal and not an alarm
condition.
• If the Auralis pump has an internal fault, the
Service indicator shows yellow, the front panel
alarm indicators show yellow and there is an
audible alarm. The Service indicator may take
up to 22 minutes to show. The alarm can be
muted (see Mute alarm button).
Low battery indicator
The single (top) indicator (see Figure 12) is shown
and an audible alarm sounds when there is a
minimum of 1 hour battery life left. The alarm can be
muted (see Mute alarm button). When there is less
than 15 minutes of battery life left, the Auralis pump
enters the Reactive CLP mode. At this point, the
audible alarm cannot be muted.
Battery charge indicators
The three indicators (blue) (see Figure 13) indicate
the charge level of the battery. When an indicator
LVÀDVKLQJLWLQGLFDWHVWKDWFKDUJLQJLVWDNLQJSODFH
When charged the indicator is constantly on. Each
indicator represents a 1/3 of the full charge of the
battery.
Low pressure
indicator
The Low Pressure indicator (see Figure 14) is shown
within 30 minutes if the Auralis pump detects low
pressure within the Auralis mattress or Auralis seat
cushion. There is a single warning tone.
If the mattress or seat cushion is disconnected while
Continue on the next page.
Содержание Auralis
Страница 1: ...INSTRUCTIONS FOR USE Auralis 04 AI 00EN_03 10 2020 Alternating pressure system ...
Страница 11: ...HOME CARE 11 Intentionally left blank ...
Страница 51: ...TROUBLESHOOTING AND ALARMS 51 Intentionally left blank ...
Страница 66: ...PARTS AND ACCESSORIES 66 Intentionally left blank ...