CLEANING AND DISINFECTION INSTRUCTIONS
40
Cleaning and disinfection instructions
The Auralis system should be routinely cleaned and disinfected after each patient and at regular intervals while
in use, as local practice for all reusable medical devices.
If there are any questions regarding cleaning and disinfecting the equipment contact Arjo Customer Service.
Make sure the Safety Data Sheets (SDS) are available for the disinfectants being used.
WARNING
Always read the instructions for use and the safety data sheet of the disinfectant.
WARNING
To avoid electrical shock, always disconnect the Auralis pump from the power source before
cleaning and inspecting.
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions
for Use.
WARNING
To avoid eye or skin irritation, never disinfect in the presence of a patient.
CAUTION
To avoid equipment damage:
• Do not use Phenol-based solutions or abrasive compounds or pads during the decontamination
process as these damage the surface coating.
• Do not spray cleaning solutions directly onto the pump.
• Do not autoclave or boil any part of the system. Avoid immersing electrical parts in water.