HYDROSENSE HS2410
Revision 5.0
P a g e 5 | 27
SPECIFICATIONS
Specifications are subject to change without notice:
Specification
Details
Power Input:
(Specify at the time of
order)
100 -130VAC*
OR
207 - 253VAC* 50/60 Hz, 200 VA max,
OPTIONAL: 24 VDC @ 3.5 A max
*Operational Range
User Interface:
Display
Four line LCD with simultaneous display in PPM, Cal Location, current
temperature, and bar graph
Communication Interface:
RS-485 Modbus, optional HART or Fieldbus Foundation module for uni-
directional communication of ppm.
Relays / Analog Outputs:
Relay Outputs
4 independent SPDT, 10 amp (7 amp if used in Class I, Div 2 applications),
dry contacts with LED panel indication
2 setpoint alarm relays (R1 & R2): user settable alarm points and delay time
(0-99 seconds delay on)
1 maintenance alarm relay (R3)
1 Lamp or controller failure alarm relay (R4)
mA Signal Output
4-20 mA DC, 900 Ohms, isolated, field scalable
Instrument Performance:
Measuring Range
0 - 5000 ppm Hydrocarbon in Water
Instrument Accuracy
± 0.1 PPM
Process Accuracy
+/- 1.0 ppm typical
The process accuracy is reflected by the site calibration to a known
hydrocarbon concentration and a stable background water. Changes in the
hydrocarbon make-up and variations in the process may affect the
instrument output.
Sensitivity
145 PPB (diesel reference) & 463 PPB (crude oil reference)
Calibration
A library of up to 10 calibrations with up to five concentration entry points per
calibration to maximize accuracy
Signal Filtering
20-1000 samples/average
Environmental:
Ambient Temperature
5-55
C
Protect from direct sun or rain. Instrument shelter or indoor use is
recommended. Higher temperatures may be accommodated with air
conditioning.
Relative humidity
Up to 90% (non-condensing)
Process Requirement:
Process Sample
Temperature
0-40
C without cooler; above 40
C with cooler.
Inlet Flow Rate
Minimum: 1.0 L/M (continuous and stable)
Optimum: 3.0 to 5.0 L/M (continuous and stable)
Inlet pressure
Minimum 2 psi, maximum 1000 psi,
*
minimum 20 psi when equipped with
optional cooler.
*For safety reasons a reducing valve should be used for pressure above 100 psi.