6.2.4 LIGHT REFERENCE AND TEMPERATURE SETTING
With automatic calibration, the light reference and temperature are automatically recorded (REF @
CAL, TEMP @ CAL). The procedure to change these values is as follows:
Press the <CALIB> key, enter the password at the prompt, then press <3> to ManRef.
Enter the new REF @ CAL value at grab sample time, then press <ENTER>.
Enter the new TEMP @ CAL value, then Press <ENTER>.
Press <DISPLAY> to main display menu.
6.2.5 CHANGE CALIBRATION LOCATION
There are 10 calibration locations in Hydrosense 2410 (Cal 0 – Cal 9). Each location can have up to 5
points to draw a calibration curve. The procedure to change calibration location is as follows:
Press the <CALIB> key, enter the password at the prompt, then press <5> to Sel Cal.
Enter the location number (0-9), then press <DISPLAY> to main display menu.