p27
Teleloading of programming files
Access to the submenu teleloading of programming
files
Alternating with
CONS
: reading of the programming file
stored in the µconsole
Transfer of programming
Indicates a writing of the programming from the TPI-SI to the
µconsole (saving of a file)
Indicates a teleloading of a programming file from the µconsole
to the TPI-SI (restoring of a file)
Programming file incorrect
The file is not made for instruments with the same
options as the instrument currently connected
LOAD
READ
COPY
µ
Co
ERR
ERR
PROG
OPT
ERR
tYPE
CONS
µ
Co
INF
SUP
CLrM
CLrA
Further functions
Display of the minimum value
Display of the maximum value
Deleting of the min. and max.
Deleting of the recorded alarms
ERR1
OPEN
2000
OL
- - - -
ERxx
Error messages
Value set out of range
Sensor rupture
Blinking measure : measure in overstepping
Displayable value overload
Upper or lower input electrical overstepping
Self-diagnosis error
The file contained in the µconsole is not for the
same type of instrument as the instrument currently connected.
«
Address
200
202
204
208
210
212
290
291
Value of the analog output
(n°1 on TPI-SI 440/441)
in µA (mA output)
Minimum value of the
displayed value
Maximum value of the
displayed value
Direct measure
Value of the analog output
(n°2 on TPI-SI 440/441)
in µA (mA output)
measure
State of relay 1
State of relay 2
Format
double integer
double integer
double integer
double integer
double integer
double integer
integer
integer
nbr of words
2
2
2
2
2
2
1
1
9. ANNEXE: MODBUS
9.1 Table of the modbus addresses
•
Direct measure:
Value without scale factor for inputs 100 mV, 1V, 10V, 20 mA :
- in mV for the 10V input
- in 1/10th of mV for the 1V input
- in µA for the mA input
- in 1/100th of mV for the mV input
Value of the resistance in 1/100
th
of
Ω
for Ni100 and Pt100
Value of the temperature of the hot sensor in 1/10
th
of degree for *
∆
Pt100
Value of the resistance
- in 1/100
th
Ω
for the resistance input from 0 to 400
Ω
- in 1/10
th
Ω
for the resistance input from 0 to 2000
Ω
- in 1/10
th
Ω
for the resistance input from 0 to 8000
Ω
Value in µV for the thermocouple input*
Value in 1/100
th
of % for the potentiometer input
* not available in the fast version
•
State of the relays
:
bit 15
bit bit
bit
1 : Led lit
1 : Led blinking : mode
recording, and measu-
re not in alarm
1 : Relay ON
Alarm
recorded