p20
Note
: The instrument no longer measures during the simulation. The
analog output and the relays will react according to the entered display.
If alarm messages have been programmed, they may appear during the
simulation.
SIMU
SIMU
Value of the
display
if
M
Press Menu
to revert to
measure
display.
4.11 Functions which can be accessed from the main
menu
4.11.1 Display simulation
(accessible according to the programmed access code and if
option relays or analog output)
The display can be simulated with the converter in order to validate the
configuration of the analog output and the relay outputs in the
installation.
b/ 2 analog outputs of the same type
(2 current outputs)
a/ 1 analog output
(current output)
Entering of a new
display value
GENE
GENE
Value of the
output
if
Enter the value
to be injected
M
Press Menu
to revert to
measure dis-
play.
Note
: The instrument will carry on measuring during the simulation. Only
the analog output no longer reacts to the measure.
GENE
GEN 1
Value of the
output
if
Enter the value
to be injected
M
Press Menu
to revert to
measure
display.
GEN 2
Value of the
output
if
Enter the value
to be injected
M
Press Menu
to revert to
measure
display.
The value of the analog output is generated simultaneously on the two
outputs.
Note
: The instrument will carry on measuring during the simulation. Only
the 2 analog outputs no longer react to the measure.
(accessible according to the programmed access code and if
option analog output)
4.11.2 Simulation of the analog outputs
(mode generator)
TPI-SI 41
TPI-SI 40
TPI-SI 441
TPI-SI 440
Содержание TPI-SI 40 Series