OPERATING INSTRUCTIONS 

ORIX 70 

 

ORDINARY AND SPECIAL MAINTENANCE  

 
 

Periodically and at least once every year, clean the coaxial electric contacts of connection between 
the pivot of the generator and the supporting bush by qualified personnel. 
 
It  is  essential  to  carefully  remove  any  possible  formation  of  brass  powder  due  to  the  use  of  the 
equipment.  
 
Every  year,  or  when  it  is  necessary,  verify  the  lubrication  of  the  moving  components  and  of  the 
joints by qualified personnel. 

 

If  the  movement  of  setting  should  be  too  much  hard,  do  not  insist  to  avoid  any  problem  of 
damaging.  
 
Any  servicing  should  be  carried  out  by  qualified  personnel  authorized  by  ARDET  Dental  & 
Medical Devices s.r.l.  
 
Every twelve months check:  

 

a)

 

That the nameplates are complete and legible 

b)

 

That the ABS coverings are upright  

c)

 

That the x-ray control board is upright.  

d)

 

The integrity of the coiled cord of the x-ray remote control (if installed) 

e)

 

The setting of the balancing spring 

f)

 

Check the insulation of the feeding cable 

 

Every year check the parameters of operation (high tension, exposure time, tube current).  
 

 

CLEANING AND DISINFECTION  

 

 
Products like polish or car-wax, normally available on the market, are suitable for a normal cleaning 
of finished surfaces. 
 
Any oil residue can be wiped off by using alcohol and passed again with a light layer of  polish or 
car-wax.  
 
Do not insist on passing alcohol on nameplates and plastic parts.  

 

For  any  disinfection  (The  rubber  protection  of  the  collimator,  which  is  in  contact  with  the  skin, 
etc.),  use  a  cold  system  of  disinfection  normally  used  in  surgery  such  as  a  solution  of  quaternary 
ammonium  /  glutamate  aldehyde  with  a  maximum  concentration  of 

2%, 

or  any  other  cooling 

disinfection systems registered by the Ministry of Health. . 

CAUTION!      

   CLEANING PRODUCTS

 

Test the products on small parts not visible.  

 

Содержание ORIX 70

Страница 1: ...l x ray control board SELEMATIC pag 15 USER S MANUAL Instructions for use SELEMATIC pag 18 Programmed exposure times pag 21 Radiographic techniques Bisector technique pag 22 Radiographic techniques Lo...

Страница 2: ...r and pantograph arm pag 45 Friction adjustment pag 47 Double jointed spring pantograph arm pag 48 Technical checkings pag 50 Possible troubles and their solutions pag 51 Spare parts pag 53 X ray cont...

Страница 3: ...e safety codes in force The electrical security of the unit is granted only if it is connected to a regular grounding conductor Any servicing should be carried out by ARDET s Dental Medical Devices s...

Страница 4: ...ices s r l if they have not been previously agreed and confirmed in writing will be returned to the sender Guarantee expires in case of Accidental breakage Carelessness The electrical wiring does not...

Страница 5: ...per evitare che la popolazione sia esposta al rischio di Radiazioni Ionizzanti superiori alle dosi massime ammissibili art 95 DPR 185 64 3 Assicurare la sorveglianza fisica della protezione a mezzo d...

Страница 6: ...erforming any operations Type of protection against direct and indirect contacts class II Degree of protection against direct and indirect contacts unit with applied part type B CAUTION Consult the te...

Страница 7: ...on level to which the patients and operating personnel are exposed c During x ray emissions only the operating personnel and the patient must be present d In case it should be necessary to help patien...

Страница 8: ...t back to the manufacturer or given to a company authorized to dispose of mineral oil and the special by products REFERENCE RULES The ORIX 70 dental x ray equipments conform with the following rules C...

Страница 9: ...ING INSTRUCTIONS ORIX 70 DESCRIPTION OF THE X RAY EQUIPMENT DESCRIPTION 1 X RAY GENERATOR TUBE HEAD 2 CENTERING CONE FSD 20 CM 3 COUNTER BALANCED ARM 4 X RAY CONTROL 5 REMOTE CONTROL 6 BASE WITH WHEEL...

Страница 10: ...x 0 8 I E C Tolerance on the reference axis of focal spot position 1 Inherent filtration Equivalent to mm 0 5 AL Total filtration or at mm 2 5 AL Limiting device cylindrical type Irradiated field dia...

Страница 11: ...1 digit Calculation of exposure time On reaching 60 of the cycle The exposure time is calculated by measuring the radiation time counting the number of cycles and half cycles where a significant quan...

Страница 12: ...manufacturer 3 Balanced spring arm plate Pantograph arm located in the back side of the holding hub of the generator It indicates serial number manufacturer 4 X ray generator plate located above the u...

Страница 13: ...Pag 13 OPERATING INSTRUCTIONS ORIX 70 1 2 3...

Страница 14: ...Pag 14 OPERATING INSTRUCTIONS ORIX 70 4 5...

Страница 15: ...DIGITAL X RAY CONTROL BOARD SELEMATIC X ray control controlled by a microprocessor for the setting management and activation of the exposure times for dental x ray systems CAUTION Carefully read the i...

Страница 16: ...time is 3 seconds and following b display of the set exposure time by default Film sensitivity d 0 05 seconds patient medium incisor maxillary arcade 1 digit alphanumeric display to display the film s...

Страница 17: ...number 6 12 Luminous display of the chosen tooth 7 3 digit display showing pre heating TIME exposure TIME selected 8 selector key and luminous display of the selected GENERATOR 9 Symbol and luminous d...

Страница 18: ...ch green light placed on the bottom of the x ray control right side ref n 1 2 Select the sensitivity of the radiographic film by pushing the button sequentially ref n 3 until the sensitivity of the fi...

Страница 19: ...key ref n 8 b Select the type of tooth ref n 6 for mandibular arcade and ref n 12 for maxillary arcade using the left and right arrows ref n 5 c Select the type of patient large medium small using th...

Страница 20: ...he following error codes are displayed E01 when for the almost immediate release of the x ray key the exposure time is equal or less than the minimum time requested by the filament of the x ray tube t...

Страница 21: ...0 0 40 0 40 LARGE PATIENT 0 47 0 40 0 54 0 80 SENSITIVITY 3 0 16 0 13 0 13 0 13 SMALL PATIENT 0 16 0 13 0 18 0 27 0 24 0 20 0 20 0 20 MEDIUM PATIENT 0 24 0 20 0 27 0 41 0 32 0 27 0 27 0 27 LARGE PATIE...

Страница 22: ...e used for each area The incidence angle of the main beam can be determined by using the graduated scale located on the bracket of the x ray tube Ref A Direction of the horizontal incidence angle of t...

Страница 23: ...Pag 23 OPERATING INSTRUCTIONS ORIX 70 BISECTOR TECHNIQUE MANDIBLE INCISORS 15 CANINES 15 PEMOLARS 10 MOLARS 5 BISECTOR TECHNIQUE OCCLUSIVE JAW 65 PREMOLARS 10 INTERPROXIMAL POSITION BITE WING...

Страница 24: ...s Occlusive axis parallel to the side of the film Reduction of x ray dose for a smaller divergence of the x ray beam Using the holder the x ray film is correctly and firmly positioned The disadvantage...

Страница 25: ...f the film does not be parallel to median sagittal plane c Place the film parallel to the major axis of the molars and parallel to the oral surfaces Third upper molar a Place the film diagonally throu...

Страница 26: ...ove and parallel to the occlusive plane c With the film in contact with the soft tissue place the film parallel to the long axis of the molars Third lower moral a The anterior side of the film must be...

Страница 27: ...t incorrect bath dilution 4 SHORT TRANSPARENCY cause insufficient fixing period bath exhausted 5 LIGHT ARCHED PORTION cause film partially exposed primary beam badly centered on the patient 6 STRAIGHT...

Страница 28: ...cause wrong manipulation scratches claws bending of x ray film before exposure 19 DOUBLE IMAGE cause movement the patient has moved 20 OVER EXPOSED PROXIMAL AREAS cause wrong set main beam not orthogo...

Страница 29: ...fixing detachment of the emulsion from the holder STORAGE OF THE FILMS It is suggested to keep a minimum stock sufficient for a period of 30 days max When using the films favour the older ones Packin...

Страница 30: ...rovided with a wash basin with running water The developer vertical type must be placed near the washbasin and ergonomically located for the operator CAUTION IN ORDER TO OBTAIN GOOD QUALITY IMAGES THE...

Страница 31: ...veloper is not being used it should always be kept covered to prevent oxidation of the developing baths If necessary restore chemicals level by introducing marble made of glass Frequently check chemic...

Страница 32: ...cessary for film transparency When fixing is over wash the film in running water for about 10 and let it dry very well before filing it EXAMPLE OF MANUAL TREATMENT DEVELOPING WASHING FIXING WASHING Te...

Страница 33: ...the x ray control board is upright d The integrity of the coiled cord of the x ray remote control if installed e The setting of the balancing spring f Check the insulation of the feeding cable Every y...

Страница 34: ...CAUTION Never use the holding bush for the pantograph arm nor the collimator to place the arm If necessary to carry out small movements till reach a correct position in order to carry out the x ray e...

Страница 35: ...l provides the equipment ORIX 70 with a digital x ray control model SELEMATIC The equipment disassembled in different parts is normally forwarded in an one card box packing CAUTATION The counter balan...

Страница 36: ...ross weight Kg 60 000 CAUTION Compressed springs SHIPMENT AND STORAGE Instruction concerning shipment and storage are found on the outside of the box These must be strictly observed 1 TRANSPORT AND ST...

Страница 37: ...RIPTION OF THE EQUIPMENT Model stand mobile double arm net weight Kg 50 000 N 1 X RAY GENERATOR Kg 9 500 N 1 CYLINDRICAL COLLIMATOR FSD 20 cm Kg 0 500 N 1 COUNTERBALANCED SPRING ARM Kg 10 000 N 1 COLU...

Страница 38: ...the control unit is facing the opposite side to the two legs of the base 4 Block the column on the central base tightening the nuts with the wrench 5 Place the assembled column plate on the wheels and...

Страница 39: ...8 Insert the threaded bar into the appropriate holes placed laterally on the column Proceed with the assembling and tighten strongly the two black bakelite handles 9 Go ahead with the tubehead mountig...

Страница 40: ...MAGNETO THERMAL SWITCH of 250V 10A with interruption ability of at least 6000 mA and with a distance between the conductors of at least 3 mm as requested by the norms The colours of the three wires li...

Страница 41: ...RUCTIONS ORIX 70 GENERAL DIAGRAM ELECTRICAL CONNECTIONS This connection allows the ignition of the ready warning light in a remote position from the x ray equipment when the main switch of the x ray c...

Страница 42: ...Check that all the functions relative to the setting of technical data are operating correctly a generator setting b film sensitivity setting c patient setting d tooth setting e ENTER setting f Pause...

Страница 43: ...from the bipolar switch are already connected to the terminals N 1 3 of the M2 terminal board The terminals of the power cable are connected with the M1 terminal board and locked with security bands...

Страница 44: ...Pag 44 OPERATING INSTRUCTIONS ORIX 70 RX TUBEHEADS CONNECTION CONNECTION External x ray control GENERATOR GENERATOR N TERMINAL N RX 1 7 10 RX 2 8 12 RX 3 9 14...

Страница 45: ...probable remains of packing material DESCRIPTION 1 X ray generator 2 Cylindrical centering device 3 Lateral bracket 4 Pivot with contact Check if the clutching pivot located on the lateral bracket is...

Страница 46: ...covered 4 Remove the brass wedge from its holding 5 Holding the x ray generator insert the pivot inside the bush of the spring arm Push upwards Replace the wedge in its place and check the perfect anc...

Страница 47: ...ray generator rotation on its fork removing the lateral plastic cap by means of a 5 mm screwdriver levering between the plastic cap and the fork and act on the 4 fixing screws of the brass ring Harde...

Страница 48: ...ing an angle of 90 with side B 2 Remove the locking cap 3 Insert inside the arm the tube key M13 provided and push it softly till the end 4 Hook the nut M 8 5 TURN to the RIGHT to increase load to the...

Страница 49: ...13 provided and push it softly till the end 4 Hook the nut M 8 5 TURN to the RIGHT to increase load to the LEFT to decrase load 6 Slip out the tube key M13 and check the counterbalancing of side B 7 I...

Страница 50: ...Pag 50 OPERATING INSTRUCTIONS ORIX 70 TECHNICAL CHECKINGS CAUTION...

Страница 51: ...sing little by little the exposure time 0 05 0 1 0 5 1 1 32 This operation is useful to check the perfect working of the timer and in the same time for the conditioning of the x ray tube before its no...

Страница 52: ...the audible signal is not given out Audible warning device is fulty Replace the power board During x ray emission fuse F1 T 6 3 A burns The connecting wire between timer and x ray equipment are in sho...

Страница 53: ...ses Apparently the x ray equipment works perfectly but the x ray images are not satisfactory in quality and detail few penetration and darkness Power supply less than 10 of the nominal value Check the...

Страница 54: ...tickers 17 1 3500038 Rubber ring for centering cone 18 2 3500042 Handles for column 19 1 3500652 Electronic boards for timer SELEMATIC 20 1 3500085 Graduated scale sticker 21 1 3500071 ABS covers for...

Страница 55: ...Pag 55 OPERATING INSTRUCTIONS ORIX 70 GENERAL ELECTRICAL SCHEME...

Страница 56: ...Pag 56 OPERATING INSTRUCTIONS ORIX 70...