OPERATING INSTRUCTIONS
ORIX 70
TECHNICAL DATA SELEMATIC
Power
230 VAC
Frequency
50 Hz
Minimum selectable time
0.03 seconds
Maximum selectable time
1,32 seconds
Exposure switch
On “DEAD MAN”
RESET to the last exposure value
Automatic
Error signal
In the case of immediate or anticipated release of the
x-ray key
“Sleeping mode” – Duty cycle
1:30
Precision
+/- 1 digit
Calculation of exposure time
On reaching 60% of the cycle
The exposure time is calculated by measuring the
radiation time counting the number of cycles and
half-cycles where a significant quantity of radiation
is emitted. (CEI 60601-2-7).
FUSE TABLE SELEMATIC
ID
VALUE
PROTECTION
POSITION
FU1 T.6.3 A
230 VAC line
Power card
FU2 F.0.16 A
Primary transformer line
Power card
FU3 0,40 A
Secondary transformer timer
Power card
FU4 0,25A
Secondary transformer timer
Power card
FU5 ¼ A
Pico fuse time
Power card
Fuse abbreviations
T. = delayed fuse
F. = quick fuse