Service with Ep88 for - 83°C degree boxes:

Note that the EP-88 gas can only be filled from a bottle containing a measured amount. Normal procedures 
used in conjunction with other refrigerants cannot be used.

Furthermore note that the freezing capability is very dependent upon filling the entire contents of the bottle 
into the system. Freezing capability is also very dependent upon the temperature of the EP88 and whether 
it has been still for a long period of time. The temperature of the service bottle must be between 8 and 14 
degrees, and the contents must be mixed by shaking the bottle 10-15 times.
In order to carry out the service procedure correctly, it is necessary to use a capillary pipe with mounting 
screws between the bottle and the filling station instead of a normal filling hose. The service technician may 
do this by using the mounting screws of a normal filling hose onto a capillary pipe.

Filling procedure:

Empty the system and evacuate to 02 mbar. Evacuation must be performed at two places in the 

system; both from the filling pipe and from the pipe mounted on the pressure side of the compressor.

Mount the capillary pipe with mounting screws between the filling station and the service bottle.

Unlike with other refrigerants, the service bottle must be turned tap-down.

Commence filling.

The compressor will start after approx. 10-15 seconds of filling.

The bottle is empty when the pressure falls to 1bar and must be shut immediately. Immediate 

shutting of the bottle is very important to prevent the fluid being pushed back into the bottle.

Close the system.

Relocating or moving the freezer:

Never leave the freezer to others without instructing these people in its use and the safety regulations 
described in the operating instructions. Always let an authorised electrician install the freezer. (Remember to 
wait for one hour before restarting it).

If the freezer and its electrical parts are reconstructed, the guarantee and safety regulations shall cease to 
apply.

If the freezer does not work: call an authorised servicing engineer!

Contact the distributor or an authorised service repair shop!

Содержание SUF 100

Страница 1: ...OPERATING INSTRUCTIONS SUF MODELS...

Страница 2: ...5 Control panel 6 Main functions 7 How to lock the keyboard 8 Parameters 9 Alarms 10 Alarm relay management 10 Digital input 10 Start defrost 12 Alarm recovery 12 Connections 13 Default setting value...

Страница 3: ...s WARNING Keep ventilation openings that are placed in the appliance enclosure or in the built in structure clear of obstruction WARNING Do not use mechanical devices or other means to accelerate the...

Страница 4: ...re is too high there is a risk that the freezer will not operate correctly It is recommended that the room is ventilated so that the room temperature does not exceed 77 F at most Class N 3 Place the f...

Страница 5: ...l parts it may cause short circuiting 6 In the event of functional or electrical irregularities please contact an authorised servicing engineer Before commissioning Clean the freezer both inside and o...

Страница 6: ...emperature display USE OF LEDS Each LED function is described in the following table 1 Press and release the key 2 The Lo message will be displayed followed by the minimum temperature recorded 3 By pr...

Страница 7: ...EFROST Push the DEF key for more than 2 seconds and a manual defrost will start HOW TO CHANGE A PARAMETER VALUE To change the parameter s value operate as follows 1 Enter the Programming mode by press...

Страница 8: ...sage will be displayed and the keyboard will be locked At this point it will be possible only to see the set point or the MAX o Min temperature stored 3 If a key is pressed more than 3s the POF messag...

Страница 9: ...n Compressor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty p...

Страница 10: ...r failure y the alarm relay is activated if a temperature alarm happens during a power failure n the alarm relay is never activated during a power failure ALF Alarm relay activation for all the alarms...

Страница 11: ...1F parameter DOOR SWITCH INPUT i1F dor It signals the door status and the corresponding relay output status through the odc parameter no Fan normal any change CPr F_C Compressor OFF Since the door is...

Страница 12: ...unction is enabled until the digital input is activated DIGITAL INPUTS POLARITY The digital input polarity depends on the i1P parameter i1P CL the input is activated by closing the contact i1P OP the...

Страница 13: ...13 CONNECTIONS XR30CX 20A COMPRESSOR 12Vac dc supply connect to the terminals 7 and 8 24Vac dc supply connect to the terminals 7 and 8 120Vac supply connect to the terminals 7 and 8...

Страница 14: ...14 82 82 95 60 DEFAULT SETTING VALUES...

Страница 15: ...te to 02 mbar Evacuation must be performed at two places in the system both from the filling pipe and from the pipe mounted on the pressure side of the compressor Mount the capillary pipe with mountin...

Страница 16: ...oose fold which is up is cut off Beware not to damage the rest of the gasket 3 The new gasket with the removed fold is placed on the lid 4 The upper part of the gasket is twisted outwards and the scre...

Страница 17: ...side the box and if it is not removed it will cause the lid to open 6 Use authorised cleaning products The use of alcohol dissolvents and alkaline cleaning agents is prohibited 7 Clean the vent holes...

Страница 18: ...18 for SUF 100 External alarm Earth connection in compressor compartment External alarm Sensor Battery 6V Sensor Battery Terminal blocks...

Страница 19: ...19 for SUF 500 External alarm Earth connection in compressor compartment External alarm Sensor Battery 6V Terminal blocks Sensor Battery Starting relay...

Страница 20: ...LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 5080100 04 GB WWW ARCTIKO INT COM WWW ARCTIKO INT COM...

Отзывы: