ALARMS

ALARM RELAY MANAGEMENT

DIGITAL INPUT

ALC Temperature alarms configuration:

 (Ab; rE)

Ab= absolute temperature: alarm temperature is given by the ALL or ALU values. rE = temperature 
alarms are referred to the set point. Temperature alarm is enabled when the temperature exceeds the 
“SET+ALU” or “SET-ALL” values.

ALU MAXIMUM temperature alarm:

 (SET÷150°C; SET÷302°F) when this temperature is reached the 

alarm is enabled, after the “ALd” delay time.

ALL Minimum temperature alarm:

 (-100.0 ÷ SET°C; -148÷302°F) when this temperature is reached the 

alarm is enabled, after the “ALd” delay time.

AFH Differential for temperature alarm recovery:

 (0,1÷25,5°C; 1÷45°F) Intervention differential for 

recovery of temperature alarm.

Ald Temperature alarm delay:

 (0÷255 min) time interval between the detection of an alarm condition and 

alarm signalling.

dAo Exclusion of temperature alarm at startup:

 (from 0.0 min to 23.5h) time interval between the 

detection of the temperature alarm condition after instrument power on and alarm signalling.

tbA Alarm relay silencing (with oA1=ALr):

 (n= silencing disabled: alarm relay stays on till alarm 

condition lasts, y =silencing enabled: alarm relay is switched OFF by pressing a key during an alarm).

Aro Alarm relay activation with power failure:

y = the alarm relay is activated if a temperature alarm happens during a power failure
n = the alarm relay is never activated during a power failure

ALF Alarm relay activation for all the alarms:

y = the alarm relay is activated for all the alarms
n = the alarm relay is activated only in case temperature alarms and regulation probe failure.

bon Time of buzzer restart after muting, in case of alarm duration:

 (0÷30min; with 0 the buzzer is 

always off after muting)

AoP Alarm relay polarity:

 it set if the alarm relay is open or closed when an alarm happens.

CL= terminals 1-2 closed during an alarm; oP = terminals 1-2 open during an alarm

i1P Digital input polarity:

 oP: the digital input is activated by opening the contact; CL: the digital input is 

activated by closing the contact.

i1F

Digital input configuration:

 EAL = external alarm: “EA” message is displayed; bAL = serious alarm 

“CA” message is displayed. PAL = pressure switch alarm, “CA” message is displayed; dor = door 
switch function; dEF = activation of a defrost cycle; AUS =to switch on the second relay if oA1 = AUS; 
Htr = kind of action inversion (cooling – heating); FAn = not set it; ES = Energy saving.

did:

(0÷255 min) 

with i1F= EAL or i1F = bAL digital input alarm delay:

 delay between the detection of 

the external alarm condition and its signalling. 

with i1F= dor: door open signalling delay

 

with i1F = PAL: time for pressure switch function:

 time interval to calculate the number of the 

pressure switch activation.

Содержание SUF 100

Страница 1: ...OPERATING INSTRUCTIONS SUF MODELS...

Страница 2: ...5 Control panel 6 Main functions 7 How to lock the keyboard 8 Parameters 9 Alarms 10 Alarm relay management 10 Digital input 10 Start defrost 12 Alarm recovery 12 Connections 13 Default setting value...

Страница 3: ...s WARNING Keep ventilation openings that are placed in the appliance enclosure or in the built in structure clear of obstruction WARNING Do not use mechanical devices or other means to accelerate the...

Страница 4: ...re is too high there is a risk that the freezer will not operate correctly It is recommended that the room is ventilated so that the room temperature does not exceed 77 F at most Class N 3 Place the f...

Страница 5: ...l parts it may cause short circuiting 6 In the event of functional or electrical irregularities please contact an authorised servicing engineer Before commissioning Clean the freezer both inside and o...

Страница 6: ...emperature display USE OF LEDS Each LED function is described in the following table 1 Press and release the key 2 The Lo message will be displayed followed by the minimum temperature recorded 3 By pr...

Страница 7: ...EFROST Push the DEF key for more than 2 seconds and a manual defrost will start HOW TO CHANGE A PARAMETER VALUE To change the parameter s value operate as follows 1 Enter the Programming mode by press...

Страница 8: ...sage will be displayed and the keyboard will be locked At this point it will be possible only to see the set point or the MAX o Min temperature stored 3 If a key is pressed more than 3s the POF messag...

Страница 9: ...n Compressor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty p...

Страница 10: ...r failure y the alarm relay is activated if a temperature alarm happens during a power failure n the alarm relay is never activated during a power failure ALF Alarm relay activation for all the alarms...

Страница 11: ...1F parameter DOOR SWITCH INPUT i1F dor It signals the door status and the corresponding relay output status through the odc parameter no Fan normal any change CPr F_C Compressor OFF Since the door is...

Страница 12: ...unction is enabled until the digital input is activated DIGITAL INPUTS POLARITY The digital input polarity depends on the i1P parameter i1P CL the input is activated by closing the contact i1P OP the...

Страница 13: ...13 CONNECTIONS XR30CX 20A COMPRESSOR 12Vac dc supply connect to the terminals 7 and 8 24Vac dc supply connect to the terminals 7 and 8 120Vac supply connect to the terminals 7 and 8...

Страница 14: ...14 82 82 95 60 DEFAULT SETTING VALUES...

Страница 15: ...te to 02 mbar Evacuation must be performed at two places in the system both from the filling pipe and from the pipe mounted on the pressure side of the compressor Mount the capillary pipe with mountin...

Страница 16: ...oose fold which is up is cut off Beware not to damage the rest of the gasket 3 The new gasket with the removed fold is placed on the lid 4 The upper part of the gasket is twisted outwards and the scre...

Страница 17: ...side the box and if it is not removed it will cause the lid to open 6 Use authorised cleaning products The use of alcohol dissolvents and alkaline cleaning agents is prohibited 7 Clean the vent holes...

Страница 18: ...18 for SUF 100 External alarm Earth connection in compressor compartment External alarm Sensor Battery 6V Sensor Battery Terminal blocks...

Страница 19: ...19 for SUF 500 External alarm Earth connection in compressor compartment External alarm Sensor Battery 6V Terminal blocks Sensor Battery Starting relay...

Страница 20: ...LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 5080100 04 GB WWW ARCTIKO INT COM WWW ARCTIKO INT COM...

Отзывы: