Arctiko CRYO 170-0 Скачать руководство пользователя страница 22

BA

T

GB

Electric Wiring Diagram

P

S

W

Power Switch

T

C

Temperature Controller

 

RTD, Sensok

M

g

1

·2

Magnet Relay

R1-3

Back-up relay

C

S

Power Outlet

C

F

1

·2

Condenser Fan Moter

TR

TH1·2

Thermal Relay

COM

1·2

Compressor

T

Freezer delay timer

R

L

Freezer breakdown lamp

PS1-2

Pressure controlling switch

SV1-2

Solenoid Valve

CHG-1

Battery Charger

BAT-1

Battery(DC2. 4V

BzSW

Buzzer Switch

B

z1

DC3V Buzzer

R

S

W

Switch, Remote Control

T

P

Terminal, Output Signal

CHG-2

Battery Charger

BAT-2

BzSW

Buzzer Switch

B

z2

DC12V Buzzer

Th2

Thermostat

SVSW

Solenoid Valve Switch

S

V

Solenoid Valve, 

LN

2

Main

Alarm, Remote 

Alarm

If the 

Auto Back-up System is installed (option)

Symbol

Symbol

Symbol

Description

Description

Description

AC200V

 DC

2.4V

C

 

H

G

+

-

BA

T

BZ1

BzSW

R1

R3

R1

RSW

2

AC200V

 DC

12V

BA

T2

+

-

BzSW

R1

R3

PB2

Th2

SV3

SVSW

BZ2

PS2

PS1

SV2

SV1

F1

F2

Mg2

R3

RL

Mg1

T

TH2

TH2

TH1

TH1

95

95

95

95

96

96

98

98

R2

PT100

A

B

b

TC

R2

R1

T

AL

OUT

15

14

13

12

11

5

2

3

4

PSW

CS

E  R  S  

T

Mg1

Mg2

TH1

TH2

!"#

!"#

Содержание CRYO 170-0

Страница 1: ...OPERATING INSTRUCTIONS CRYO 170 0 230 0...

Страница 2: ...hould only be operated after reading the manual and with the user s due care and discretion Particular care should be taken not to disassemble the product Warning Ignoring this warning may cause damag...

Страница 3: ...on panel Maintenance Troubleshooting Moving and Reinstalling LN2 Back up Circuit diagram Specifications Electric Wiring Diagram Power cut Turning off the appliance for long periods Repair of scratches...

Страница 4: ...ers depending on circumstances Please be sure to keep all matters since they contain important details about safety Should the product be transferred to a new location or lent to a different user plea...

Страница 5: ...e where it may come into contact with water Failure to follow these instructions may ca use electric leakage and or electric shocks due to lower electrical isolation Do not pour or spill water on the...

Страница 6: ...cked in the appliance Warning For correct use In case of gas leakage do not touch the freezer Turn off the gas valve and ventilate the area Pulling out the power plug may cause an explosion due to sta...

Страница 7: ...n the mechanical compartment While the freezer is cooling down the block condenser gets hot because of heat radiation Heat Radiation Make sure to set the sub lid after taking out or putting in preserv...

Страница 8: ...t least 50cm between the top of the freezer and the ceiling Electric Leakage Breaker If installation in a damp area is unavoidable the use of an electric leakage breaker is strongly required Install t...

Страница 9: ...ls containers especially metallic ones or the inside of the freezer with bare or wet hands Failure to follow these instructions may lead to frostbite Make sure to wear gloves when working 1 Install th...

Страница 10: ...thertypeofvessel Donotuseanyinflammablespray suchas lacquer paint or any form of volatile substances or inflammables near the freezer Sparks discharged from the electrical components of the freezer ma...

Страница 11: ...alue PV and set value SV Changing the set value by pressing Key to set the temperature Temperature Controller Names Functions of Front Display PV SV OUT 1 2 AL 1 2 COM RDY DI MODEKey FUNCKey Key Prese...

Страница 12: ...n the alarm buzzer goes off when AL 2 lamp comes on When AL 2 lamp is on please contact the sales agency Even if only one compressor is running the temperature in the freezer is maintained at approxim...

Страница 13: ...ng the button UP or DOWN 3 Access to the SET level by pressing the MODE button 4 Scroll between the parameter in the SET level by pressing the MODE button 5 To change the parameters press the button u...

Страница 14: ...12 GB 180...

Страница 15: ...13 GB...

Страница 16: ...ver make sure that the wax does not touch any plastic parts of the appliance The above mentioned substances may damage the paintwork and plastic surfaces of the appliance Especially Alkalescent deterg...

Страница 17: ...er cut Is the power switch turned off 2 When the appliance shows poor cooling performance Does the set temperature correspond to the desired temperature Is the appliance placed out of direct sunlight...

Страница 18: ...in a condition of high humidity when the doors are opened and closed frequently or when the stored materials have high moisture levels Wipe it off with a dry cloth The front and sides of the applianc...

Страница 19: ...not lay it on its side since this may became a cause of a breakdown Power cut Limit door opening as much as possible Do not put any new materials into the freezer It will make the inside temperature h...

Страница 20: ...uipment Circuit Diagram H07RN F 5G2 5mm2 WD11 H07RN F 5G2 5mm2 WD14 Supply 3x230 400V XD11 XD14 TA14 Transformer 400 200V 5 3 10 1A D Y 3 5kVA L1 L2 L3 PE L1 L2 L3 PE gy bk bn gnye gy bk bn gnye 2U 1U...

Страница 21: ...on Freon 3 220V 50 60Hz 20A or higher Digital 300 CRYO 230 0 239 W1600 D840 H1090 W680 D500 H705 3 220V 50 60Hz 20A or higher Digital 350 Do not convert the appliance It may lead to electric shock or...

Страница 22: ...Buzzer RSW Switch Remote Control TP Terminal Output Signal CHG 2 Battery Charger BAT 2 BzSW Buzzer Switch Bz2 DC12V Buzzer Th2 Thermostat SVSW Solenoid Valve Switch SV Solenoid Valve LN2 Main Alarm R...

Страница 23: ...21 GB...

Страница 24: ...LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM WWW ARCTIKO COM 50 80 056 01...