![AR 1000S1G6C Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/ar/1000s1g6c/1000s1g6c_operating-and-service-manual_2962344036.webp)
Model 1000S1G6C
12
Rev A
2.4.2.1.1
Center Conductor Inspection
Index
Description
1
New Condition
2
Normal Use-Slight gap
3
Bent pin section-Discard connector
4
Fingers spread or collapsed-Discard connector
5
Broken pin or pin section-Discard connector
2.4.2.1.2
Center Pin Concentricity
Index
Description
1
Centered
2
Maximum allowable off-center
3
Fall-off center-Discard connector
2.4.2.1.3
Cleaning
When the connector threads of the connectors are connected and disconnected, the threads and outer conductor
mating interfaces can build up a layer of dirt, metal chips, and other foreign material. These contaminants can
degrade electrical and mechanical performance. To prevent inaccurate measurements and reduced connector
life, connector interfaces must be kept clean.