Model 500T2G8
20
Rev E
The load VSWR under power can be checked by using a power meter with suitable attenuators connected to
the RF reflected sample out port A2J3. Refer to the RF reflected sample port calibration factors on the rear of
the unit. These show the relation between the amplifier power reflected back into the output ports and the RF
reflected sample port power as a function of frequency.
Set keylock to LOCAL.
Switch on the MAIN POWER circuit breaker. The fan will operate. The front panel display will show several
identification messages and then the MENU 1 screen. The third line will indicate that the heater time delay is
active.
Allow the heater warm-up delay to expire. Line three will indicate OFF/READY.
Push S4 (MORE) three times to go to MENU 4. Verify that the heater voltage and current are near their
nominal levels. The values of these parameters at the time the TWTA left the factory are logged on the test
data sheet.
Push the OPERATE push-button. You will now see the cathode and the collector voltages rise. Verify that the
collector and cathode voltages are near nominal. The helix current should be close to the nominal value for no
RF drive. Then push MORE or EXIT to go back to MENU 1.
Set the TWTA gain to maximum. Adjust the RF generator to slowly increase the RF drive toward 0 dBm to
reach the desired FWD PWR on the display and power meter (connected to the forward sample port). The
forward power bar graph will become active, with maximum length when peak power output is achieved.
Best performance is obtained when the input RF drive is set at or just below the level which causes peak
power output. Do not set input drive above 0 dBm (Input drive above +20 dBm may damage the unit). The
reflected power level should remain below 10% of the forward power, assuming that the load is properly
matched.
An alternate procedure is to pre-set the TWTA gain to minimum, set the RF generator to 0 dBm and then
slowly increase the TWTA gain to set the desired RF output level.
Observe that the helix current is sensitive to the RF drive level of the TWT. It is at a minimum with no RF
drive. The helix currents with no drive and with rated RF output mid-band are logged on the test data sheet.
The value of the helix current is a good qualitative indicator of RF drive present.
To shut the system down, turn down the RF generator level below -50 dBm and press STANDBY. Allow the
TWTA to cool down until the TWT temperature drops below 70°C, then turn off main power
3.7 EMERGENCY BYPASS OPERATION
For reference, see schematic 10-24830-000 in Section 5.2.
The TWTA is provided with a means of operating the amplifier manually in the event that there is a failure of
the control module and it is imperative that the amplifier remain on line.
CAUTION:
Emergency bypass operation disables all microprocessor based interactive
features and possibly some diagnostic features. For this reason, the
emergency bypass mode of operation should be used only when the
control unit fails and when it is essential to remain on line.
Содержание 10006896
Страница 2: ......
Страница 4: ......
Страница 8: ...Model 500T2G8...
Страница 10: ......
Страница 18: ......
Страница 40: ...Model 500T2G8 28 Rev E...
Страница 44: ...Model 500T2G8 32 Rev E...
Страница 46: ...Model 500T2G8 34 Rev E...
Страница 50: ...Model 500T2G8 38 Rev E...
Страница 52: ...Model 500T2G8 40 Rev E...
Страница 66: ...Model 500T2G8 54 Rev E...
Страница 68: ......