ORP - Diagnostics Menu - Calibration Data 5.12
Page 42
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
ORP - Diagnostics Menu - Calibration Data 5.12
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
The Calibration Data menu is a series of read only screens which
allow the user to view the data collected during the last calibration.
RUN MODE
If the Temperature Display
Override is ON (Section 5.10),
the display will show
Press to view the first
Calibration Data screen.
1st POINT 600mV
Probe Temperature at which
calibration was performed = 19.4 C
Calibration Mode
1 point calibration
Press to view the next
Calibration Data screen.
CALIBRATION
UTILITIES
DIAGNOSTICS
SETUP
500mV 25.0C
CALIBRATION DATA
TEMP O/R ON
CAL TEMP 19.4C >
CALIBRATION DATA
CALIBRATION DATA
MODE 1PT
>
CALIBRATION DATA
CALIBRATION DATA
FIRMWARE REV.
600mV
>